Atomic absorption spectroscopy (AAS) is a spectroanalytical procedure for the quantitative measurement of chemical elements. AAS is based on the absorption of light by free metallic ions that have been atomized from a sample. An alternative technique is atomic emission spectroscopy (AES).
In analytical chemistry the technique is used for determining the concentration of a particular element (the analyte) in a sample to be analyzed. AAS can be used to determine over 70 different elements in solution, or directly in solid samples via electrothermal vaporization,[1] and is used in pharmacology, biophysics,
archaeology and toxicology research.
Atomic absorption spectrometry has many uses in different areas of chemistry such as clinical analysis of metals in biological fluids and tissues such as whole blood, plasma, urine, saliva, brain tissue, liver, hair, muscle tissue. Atomic absorption spectrometry can be used in qualitative and quantitative analysis.
Principles
The technique makes use of the atomic absorption spectrum of a sample in order to assess the concentration of specific analytes within it. It requires standards with known analyte content to establish the relation between the measured absorbance and the analyte concentration and relies therefore on the Beer–Lambert law. Analyzing Samples with Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS) measures the concentration of specific elements in a sample by analyzing their unique "fingerprint" in the form of an atomic absorption spectrum. Here's how it works:
Step 1: Sample Preparation:**
The sample is typically dissolved in a suitable solvent (acids, water) to create a liquid solution. This ensures the analytes are present as free atoms, ready for absorption.
For solid samples like ores or minerals, additional steps like grinding and digestion may be required to break down the matrix and liberate the analytes.
Step 2: Atomization:**
The prepared solution is nebulized into a fine mist and introduced into a high-temperature flame (air-acetylene or nitrous oxide-acetylene mix).
The intense heat in the flame excites the electrons in the analyte atoms, promoting them to higher energy levels.
Step 3: Absorption:**
Simultaneously, a hollow cathode lamp containing the same element as the analyte emits a specific wavelength of light that corresponds to the energy difference between the excited and ground state of the analyte atoms.
As the emitted light passes through the atomized sample, some photons are absorbed by the excited analyte atoms, causing them to return to their ground state. This absorption decreases the intensity of the light at the specific wavelength.
Step 4: Measurement and Analysis:**
The light intensity before and after passing through the sample is measured by a detector.
The difference in intensity is directly proportional to the concentration of the analyte in the sample, following the Beer-Lambert law:
* **A = εcl**, where:
* A is the absorbance measured.
* ε is the molar absorptivity (constant specific to the element and wavelength).
* c is the concentration of the analyte.
* l is the path length of the light through the sample.
Step 5: Calibration and Quantification:**
To determine the actual concentration of the analyte, the instrument is calibrated using standard solutions containing known concentrations of the element.
By comparing the measured absorbance of the sample to the calibration curve, the concentration of the analyte in the original sample can be calculated.
Feedback Mechanism:**
The measured absorbance directly provides feedback on the concentration of the analyte in the sample. This feedback loop allows the AAS to analyze various samples efficiently and determine their elemental composition with high accuracy.
In summary, AAS utilizes the unique absorption properties of elements to accurately quantify their concentration in samples. By preparing the sample, atomizing the analytes, measuring their absorption of specific light, and applying the Beer-Lambert law, this powerful technique helps us understand the elemental makeup of diverse materials across various scientific and industrial fields.
Instrumentation
In order to analyze a sample for its atomic constituents, it has to be atomized. The atomizers most commonly used nowadays are flames and electrothermal (graphite tube) atomizers. The atoms should then be irradiated by optical radiation, and the radiation source could be an element-specific line radiation source or a continuum radiation source. The radiation then passes through a monochromator in order to separate the element-specific radiation from any other radiation emitted by the radiation source, which is finally measured by a detector.
Atomizers
The used nowadays are spectroscopic flames and electrothermal atomizers. Other atomizers, such as glow-discharge atomization, hydride atomization, or cold-vapor atomization, might be used for special purposes.
Flame atomizers
The oldest and most commonly used atomizers in AAS are flames, principally the air-acetylene flame with a temperature of about 2300 °C and the nitrous oxide[4] system (N2O)-acetylene flame with a temperature of about 2700 °C. The latter flame, in addition, offers a more reducing environment, being ideally suited for analytes with high affinity to oxygen.
Liquid or dissolved samples are typically used with flame atomizers. The sample solution is aspirated by a pneumatic analytical nebulizer, transformed into an aerosol, which is introduced into a spray chamber, where it is mixed with the flame gases and conditioned in a way that only the finest aerosol droplets (< 10 μm) enter the flame. This conditioning process reduces interference, but only about 5% of the aerosolized solution reaches the flame because of it.
On top of the spray chamber is a burner head that produces a flame that is laterally long (usually 5–10 cm) and only a few mm deep. The radiation beam passes through this flame at its longest axis, and the flame gas flow-rates may be adjusted to produce the highest concentration of free atoms. The burner height may also be adjusted, so that the radiation beam passes through the zone of highest atom cloud density in the flame, resulting in the highest sensitivity.
The processes in a flame include the stages of desolvation (drying) in which the solvent is evaporated and the dry sample nano-particles remain, vaporization (transfer to the gaseous phase) in which the solid particles are converted into gaseous molecule, atomization in which the molecules are dissociated into free atoms, and ionization where (depending on the ionization potential of the analyte atoms and the energy available in a particular flame) atoms may be in part converted to gaseous ions.
Each of these stages includes the risk of interference in case the degree of phase transfer is different for the analyte in the calibration standard and in the sample. Ionization is generally undesirable, as it reduces the number of atoms that are available for measurement, i.e., the sensitivity.
In flame AAS a steady-state signal is generated during the time period when the sample is aspirated. This technique is typically used for determinations in the mg L−1 range, and may be extended down to a few μg L−1 for some elements.
Electrothermal atomizers
Electrothermal AAS (ET AAS) using graphite tube atomizers was pioneered by Boris V. L’vov at the Saint Petersburg Polytechnical Institute, Russia,[5] since the late 1950s, and investigated in parallel by Hans Massmann at the Institute of Spectrochemistry and Applied Spectroscopy (ISAS) in Dortmund, Germany.[6]
Although a wide variety of graphite tube designs have been used over the years, the dimensions nowadays are typically 20–25 mm in length and 5–6 mm inner diameter. With this technique liquid/dissolved, solid and gaseous samples may be analyzed directly. A measured volume (typically 10–50 μL) or a weighed mass (typically around 1 mg) of a solid sample are introduced into the graphite tube and subject to a temperature program. This typically consists of stages, such as drying – the solvent is evaporated; pyrolysis – the majority of the matrix constituents are removed; atomization – the analyte element is released to the gaseous phase; and cleaning – eventual residues in the graphite tube are removed at high temperature.[7]
The graphite tubes are heated via their ohmic resistance using a low-voltage high-current power supply; the temperature in the individual stages can be controlled very closely, and temperature ramps between the individual stages facilitate separation of sample components. Tubes may be heated transversely or longitudinally, where the former ones have the advantage of a more homogeneous temperature distribution over their length. The so-called stabilized temperature platform furnace (STPF) concept, proposed by Walter Slavin, based on research of Boris L’vov, makes ET AAS essentially free from interference.[citation needed] The major components of this concept are atomization of the sample from a graphite platform inserted into the graphite tube (L’vov platform) instead of from the tube wall in order to delay atomization until the gas phase in the atomizer has reached a stable temperature; use of a chemical modifier in order to stabilize the analyte to a pyrolysis temperature that is sufficient to remove the majority of the matrix components; and integration of the absorbance over the time of the transient absorption signal instead of using peak height absorbance for quantification.
In ET AAS a transient signal is generated, the area of which is directly proportional to the mass of analyte (not its concentration) introduced into the graphite tube. This technique has the advantage that any kind of sample, solid, liquid or gaseous, can be analyzed directly. Its sensitivity is 2–3 orders of magnitude higher than that of flame AAS, so that determinations in the low μg L−1 range (for a typical sample volume of 20 μL) and ng g−1 range (for a typical sample mass of 1 mg) can be carried out. It shows a very high degree of freedom from interferences, so that ET AAS might be considered the most robust technique available nowadays for the determination of trace elements in complex matrices.[citation needed]
Specialized atomization techniques
While flame and electrothermal vaporizers are the most common atomization techniques, several other atomization methods are utilized for specialized use.[8][9]
Glow-discharge atomization
A glow-discharge device (GD) serves as a versatile source, as it can simultaneously introduce and atomize the sample. The glow discharge occurs in a low-pressure argon gas atmosphere between 1 and 10 torr. In this atmosphere lies a pair of electrodes applying a DC voltage of 250 to 1000 V to break down the argon gas into positively charged ions and electrons. These ions, under the influence of the electric field, are accelerated into the cathode surface containing the sample, bombarding the sample and causing neutral sample atom ejection through the process known as sputtering. The atomic vapor produced by this discharge is composed of ions, ground state atoms, and fraction of excited atoms. When the excited atoms relax back into their ground state, a low-intensity glow is emitted, giving the technique its name.
The requirement for samples of glow discharge atomizers is that they are electrical conductors. Consequently, atomizers are most commonly used in the analysis of metals and other conducting samples. However, with proper modifications, it can be utilized to analyze liquid samples as well as nonconducting materials by mixing them with a conductor (e.g. graphite).
Hydride atomization
Hydride generation techniques are specialized in solutions of specific elements. The technique provides a means of introducing samples containing arsenic, antimony, selenium, bismuth, and lead into an atomizer in the gas phase. With these elements, hydride atomization enhances detection limits by a factor of 10 to 100 compared to alternative methods. Hydride generation occurs by adding an acidified aqueous solution of the sample to a 1% aqueous solution of sodium borohydride, all of which is contained in a glass vessel. The volatile hydride generated by the reaction that occurs is swept into the atomization chamber by an inert gas, where it undergoes decomposition. This process forms an atomized form of the analyte, which can then be measured by absorption or emission spectrometry.
Cold-vapor atomization
The cold-vapor technique is an atomization method limited only for the determination of mercury, due to it being the only metallic element to have a large vapor pressure at ambient temperature.[citation needed] Because of this, it has an important use in determining organic mercury compounds in samples and their distribution in the environment. The method initiates by converting mercury into Hg2+ by oxidation from nitric and sulfuric acids, followed by a reduction of Hg2+ with tin(II) chloride. The mercury, is then swept into a long-pass absorption tube by bubbling a stream of inert gas through the reaction mixture. The concentration is determined by measuring the absorbance of this gas at 253.7 nm. Detection limits for this technique are in the parts-per-billion range making it an excellent mercury detection atomization method.
Radiation sources
We have to distinguish between line source AAS (LS AAS) and continuum source AAS (CS AAS). In classical LS AAS, as it has been proposed by Alan Walsh,[10] the high spectral resolution required for AAS measurements is provided by the radiation source itself that emits the spectrum of the analyte in the form of lines that are narrower than the absorption lines. Continuum sources, such as deuterium lamps, are only used for background correction purposes. The advantage of this technique is that only a medium-resolution monochromator is necessary for measuring AAS; however, it has the disadvantage that usually a separate lamp is required for each element that has to be determined. In CS AAS, in contrast, a single lamp, emitting a continuum spectrum over the entire spectral range of interest is used for all elements. Obviously, a high-resolution monochromator is required for this technique, as will be discussed later.
Hollow cathode lamps
Hollow cathode lamps (HCL) are the most common radiation source in LS AAS.[citation needed] Inside the sealed lamp, filled with argon or neon gas at low pressure, is a cylindrical metal cathode containing the element of interest and an anode. A high voltage is applied across the anode and cathode, resulting in an ionization of the fill gas. The gas ions are accelerated towards the cathode and, upon impact on the cathode, sputter cathode material that is excited in the glow discharge to emit the radiation of the sputtered material, i.e., the element of interest. In the majority of cases single element lamps are used, where the cathode is pressed out of predominantly compounds of the target element. Multi-element lamps are available with combinations of compounds of the target elements pressed in the cathode. Multi element lamps produce slightly less sensitivity than single element lamps and the combinations of elements have to be selected carefully to avoid spectral interferences. Most multi-element lamps combine a handful of elements, e.g.: 2 - 8. Atomic Absorption Spectrometers can feature as few as 1-2 hollow cathode lamp positions or in automated multi-element spectrometers, a 8-12 lamp positions may be typically available.
Electrodeless discharge lamps
Electrodeless discharge lamps (EDL) contain a small quantity of the analyte as a metal or a salt in a quartz bulb together with an inert gas, typically argon gas, at low pressure. The bulb is inserted into a coil that is generating an electromagnetic radio frequency field, resulting in a low-pressure inductively coupled discharge in the lamp. The emission from an EDL is higher than that from an HCL, and the line width is generally narrower, but EDLs need a separate power supply and might need a longer time to stabilize.
Deuterium lamps
Deuterium HCL or even hydrogen HCL and deuterium discharge lamps are used in LS AAS for background correction purposes.[11] The radiation intensity emitted by these lamps decreases significantly with increasing wavelength, so that they can be only used in the wavelength range between 190 and about 320 nm.
Continuum sources
When a continuum radiation source is used for AAS, it is necessary to use a high-resolution monochromator, as will be discussed later. In addition, it is necessary that the lamp emits radiation of intensity at least an order of magnitude above that of a typical HCL over the entire wavelength range from 190 nm to 900 nm. A special high-pressure xenon short arc lamp, operating in a hot-spot mode has been developed to fulfill these requirements.
Spectrometer
As already pointed out above, there is a difference between medium-resolution spectrometers that are used for LS AAS and high-resolution spectrometers that are designed for CS AAS. The spectrometer includes the spectral sorting device (monochromator) and the detector.
Spectrometers for LS AAS
In LS AAS the high resolution that is required for the measurement of atomic absorption is provided by the narrow line emission of the radiation source, and the monochromator simply has to resolve the analytical line from other radiation emitted by the lamp.[citation needed] This can usually be accomplished with a band pass between 0.2 and 2 nm, i.e., a medium-resolution monochromator. Another feature to make LS AAS element-specific is modulation of the primary radiation and the use of a selective amplifier that is tuned to the same modulation frequency, as already postulated by Alan Walsh. This way any (unmodulated) radiation emitted for example by the atomizer can be excluded, which is imperative for LS AAS. Simple monochromators of the Littrow or (better) the Czerny-Turner design are typically used for LS AAS. Photomultiplier tubes are the most frequently used detectors in LS AAS, although solid state detectors might be preferred because of their better signal-to-noise ratio.
Spectrometers for CS AAS
When a continuum radiation source is used for AAS measurement it is indispensable to work with a high-resolution monochromator. The resolution has to be equal to or better than the half-width of an atomic absorption line (about 2 pm) in order to avoid losses of sensitivity and linearity of the calibration graph. The research with high-resolution (HR) CS AAS was pioneered by the groups of O’Haver and Harnly in the US, who also developed the (up until now) only simultaneous multi-element spectrometer for this technique. The breakthrough, however, came when the group of Becker-Ross in Berlin, Germany, built a spectrometer entirely designed for HR-CS AAS. The first commercial equipment for HR-CS AAS was introduced by Analytik Jena (Jena, Germany) at the beginning of the 21st century, based on the design proposed by Becker-Ross and Florek. These spectrometers use a compact double monochromator with a prism pre-monochromator and an echelle grating monochromator for high resolution. A linear charge-coupled device (CCD) array with 200 pixels is used as the detector. The second monochromator does not have an exit slit; hence the spectral environment at both sides of the analytical line becomes visible at high resolution. As typically only 3–5 pixels are used to measure the atomic absorption, the other pixels are available for correction purposes. One of these corrections is that for lamp flicker noise, which is independent of wavelength, resulting in measurements with very low noise level; other corrections are those for background absorption, as will be discussed later.
Background absorption and background correction
The relatively small number of atomic absorption lines (compared to atomic emission lines) and their narrow width (a few pm) make spectral overlap rare; there are only few examples known that an absorption line from one element will overlap with another.[citation needed] Molecular absorption, in contrast, is much broader, so that it is more likely that some molecular absorption band will overlap with an atomic line. This kind of absorption might be caused by un-dissociated molecules of concomitant elements of the sample or by flame gases. We have to distinguish between the spectra of di-atomic molecules, which exhibit a pronounced fine structure, and those of larger (usually tri-atomic) molecules that don't show such fine structure. Another source of background absorption, particularly in ET AAS, is scattering of the primary radiation at particles that are generated in the atomization stage, when the matrix could not be removed sufficiently in the pyrolysis stage.
All these phenomena, molecular absorption and radiation scattering, can result in artificially high absorption and an improperly high (erroneous) calculation for the concentration or mass of the analyte in the sample. There are several techniques available to correct for background absorption, and they are significantly different for LS AAS and HR-CS AAS.
Background correction techniques in LS AAS
In LS AAS background absorption can only be corrected using instrumental techniques, and all of them are based on two sequential measurements:[12] firstly, total absorption (atomic plus background), secondly, background absorption only. The difference of the two measurements gives the net atomic absorption. Because of this, and because of the use of additional devices in the spectrometer, the signal-to-noise ratio of background-corrected signals is always significantly inferior compared to uncorrected signals. It should also be pointed out that in LS AAS there is no way to correct for (the rare case of) a direct overlap of two atomic lines. In essence there are three techniques used for background correction in LS AAS:
Deuterium background correction
This is the oldest and still most commonly used technique, particularly for flame AAS. In this case, a separate source (a deuterium lamp) with broad emission is used to measure the background absorption over the entire width of the exit slit of the spectrometer. The use of a separate lamp makes this technique the least accurate one, as it cannot correct for any structured background. It also cannot be used at wavelengths above about 320 nm, as the emission intensity of the deuterium lamp becomes very weak. The use of deuterium HCL is preferable compared to an arc lamp due to the better fit of the image of the former lamp with that of the analyte HCL.
Smith-Hieftje background correction
This technique (named after their inventors) is based on the line-broadening and self-reversal of emission lines from HCL when high current is applied. Total absorption is measured with normal lamp current, i.e., with a narrow emission line, and background absorption after application of a high-current pulse with the profile of the self-reversed line, which has little emission at the original wavelength, but strong emission on both sides of the analytical line. The advantage of this technique is that only one radiation source is used; among the disadvantages are that the high-current pulses reduce lamp lifetime, and that the technique can only be used for relatively volatile elements, as only those exhibit sufficient self-reversal to avoid dramatic loss of sensitivity. Another problem is that background is not measured at the same wavelength as total absorption, making the technique unsuitable for correcting structured background.
An alternating magnetic field is applied at the atomizer (graphite furnace) to split the absorption line into three components, the π component, which remains at the same position as the original absorption line, and two σ components, which are moved to higher and lower wavelengths, respectively.[citation needed] Total absorption is measured without magnetic field and background absorption with the magnetic field on. The π component has to be removed in this case, e.g. using a polarizer, and the σ components do not overlap with the emission profile of the lamp, so that only the background absorption is measured. The advantages of this technique are that total and background absorption are measured with the same emission profile of the same lamp, so that any kind of background, including background with fine structure can be corrected accurately, unless the molecule responsible for the background is also affected by the magnetic field and using a chopper as a polariser reduces the signal to noise ratio. While the disadvantages are the increased complexity of the spectrometer and power supply needed for running the powerful magnet needed to split the absorption line.
Background correction techniques in HR-CS AAS
In HR-CS AAS background correction is carried out mathematically in the software using information from detector pixels that are not used for measuring atomic absorption; hence, in contrast to LS AAS, no additional components are required for background correction.
Background correction using correction pixels
It has already been mentioned that in HR-CS AAS lamp flicker noise is eliminated using correction pixels. In fact, any increase or decrease in radiation intensity that is observed to the same extent at all pixels chosen for correction is eliminated by the correction algorithm.[citation needed] This obviously also includes a reduction of the measured intensity due to radiation scattering or molecular absorption, which is corrected in the same way. As measurement of total and background absorption, and correction for the latter, are strictly simultaneous (in contrast to LS AAS), even the fastest changes of background absorption, as they may be observed in ET AAS, do not cause any problem. In addition, as the same algorithm is used for background correction and elimination of lamp noise, the background corrected signals show a much better signal-to-noise ratio compared to the uncorrected signals, which is also in contrast to LS AAS.
Background correction using a least-squares algorithm
The above technique can obviously not correct for a background with fine structure, as in this case the absorbance will be different at each of the correction pixels. In this case HR-CS AAS is offering the possibility to measure correction spectra of the molecule(s) that is (are) responsible for the background and store them in the computer. These spectra are then multiplied with a factor to match the intensity of the sample spectrum and subtracted pixel by pixel and spectrum by spectrum from the sample spectrum using a least-squares algorithm. This might sound complex, but first of all the number of di-atomic molecules that can exist at the temperatures of the atomizers used in AAS is relatively small, and second, the correction is performed by the computer within a few seconds. The same algorithm can actually also be used to correct for direct line overlap of two atomic absorption lines, making HR-CS AAS the only AAS technique that can correct for this kind of spectral interference.
^ abKoirtyohann, S. R. (1991). "A History of Atomic Absorption Spectrometry". Analytical Chemistry. 63 (21): 1024A–1031A. doi:10.1021/ac00021a716. ISSN0003-2700.
B. Welz, M. Sperling (1999), Atomic Absorption Spectrometry, Wiley-VCH, Weinheim, Germany, ISBN3-527-28571-7.
A. Walsh (1955), The application of atomic absorption spectra to chemical analysis, Spectrochim. Acta 7: 108–117.
J.A.C. Broekaert (1998), Analytical Atomic Spectrometry with Flames and Plasmas, 3rd Edition, Wiley-VCH, Weinheim, Germany.
B.V. L’vov (1984), Twenty-five years of furnace atomic absorption spectroscopy, Spectrochim. Acta Part B, 39: 149–157.
B.V. L’vov (2005), Fifty years of atomic absorption spectrometry; J. Anal. Chem., 60: 382–392.
H. Massmann (1968), Vergleich von Atomabsorption und Atomfluoreszenz in der Graphitküvette, Spectrochim. Acta Part B, 23: 215–226.
W. Slavin, D.C. Manning, G.R. Carnrick (1981), The stabilized temperature platform furnace, At. Spectrosc. 2: 137–145.
B. Welz, H. Becker-Ross, S. Florek, U. Heitmann (2005), High-resolution Continuum Source AAS, Wiley-VCH, Weinheim, Germany, ISBN3-527-30736-2.
H. Becker-Ross, S. Florek, U. Heitmann, R. Weisse (1996), Influence of the spectral bandwidth of the spectrometer on the sensitivity using continuum source AAS, Fresenius J. Anal. Chem. 355: 300–303.
J.M. Harnly (1986), Multi element atomic absorption with a continuum source, Anal. Chem. 58: 933A-943A.
Skoog, Douglas (2007). Principles of Instrumental Analysis (6th ed.). Canada: Thomson Brooks/Cole. ISBN0-495-01201-7.
Mohammed Abdellaoue Informasi pribadiNama lengkap Mohammed AbdellaoueTanggal lahir 23 Oktober 1985 (umur 38)Tempat lahir Oslo, NorwegiaTinggi 1,82 m (5 ft 11+1⁄2 in)Posisi bermain PenyerangInformasi klubKlub saat ini VålerengaKarier junior1997–2000 Hasle-Løren2001–2003 SkeidKarier senior*Tahun Tim Tampil (Gol)2003–2007 Skeid 86 (40)2008–2010 Vålerenga 67 (30)2010–2013 Hannover 96 80 (29)2013–2015 VfB Stuttgart 12 (1)2015 VfB Stuttgart II 4 (0)2015– ...
AlectisRentang fosil: 55–0 jtyl PreЄ Є O S D C P T J K Pg N Eocene to Present[1] Alectis alexandrina Klasifikasi ilmiah Kerajaan: Animalia Filum: Chordata Kelas: Actinopterygii Ordo: Carangiformes Famili: Carangidae Subfamili: Caranginae Genus: AlectisRafinesque, 1815 Spesies tipe Alectis ciliaris (Bloch, 1787) Spesies[2] Alectis alexandrina (É. Geoffroy Saint-Hilaire, 1817) Alectis ciliaris (Bloch, 1787) Alectis indica (Ruppell, 1830) Sinonim Gallus (Lacépède, 18...
العلاقات الفلبينية الدومينيكية الفلبين دومينيكا الفلبين دومينيكا تعديل مصدري - تعديل العلاقات الفلبينية الدومينيكية هي العلاقات الثنائية التي تجمع بين الفلبين ودومينيكا.[1][2][3][4][5] مقارنة بين البلدين هذه مقارنة عامة ومرجعية للدولتي�...
WinstonJenis produkRokok PutihPemilikJapan TobaccoProdusenPT Alam IndomegahNegaraAmerika SerikatDiluncurkan1954; 70 tahun lalu (1954)PasarIndonesiaPemilik sebelumnyaR.J. Reynolds Winston adalah merk rokok asal Amerika yang diproduksi oleh Japan Tobacco dan di Indonesia diproduksi oleh PT Alam Indomegah. Produk Winston menjadi merk unggulan dunia oleh Japan Tobacco dan diluncurkan di Indonesia pada tahun 2015, dan dulu sempat di import langsung dari Malaysia. Sejarah Winston diluncurkan p...
View of the paradis in Théâtre de Douai View of the stage from the paradis of the Théâtre du Jeu de Paume, Aix-en-Provence The gods (UK English), or sometimes paradise, is a theatrical term referring to the highest areas of a theatre such as the upper balconies. These are generally the cheapest seats; the moniker may have come from the ornately painted ceilings in older venues, often based on mythological themes.[citation needed] Similarly those seated so high up look down upon bo...
Election in Nebraska Main article: 1868 United States presidential election 1868 United States presidential election in Nebraska November 3, 1868 1872 → Nominee Ulysses S. Grant Horatio Seymour Party Republican Democratic Home state Illinois New York Running mate Schuyler Colfax Francis Preston Blair, Jr. Electoral vote 3 0 Popular vote 9,772 5,519 Percentage 63.91% 36.09% Grant 50–60% 60–70% 70–80% 80...
The Central Auditing Organization (Arabic: الجهاز المركزي للمحاسبات) is an independent auditing institution established in Egypt in 1942 as an instrument of public finance control. Originally created as the Divan of Accounting, its name was changed to Divan of Auditing in 1960, before acquiring its current designation in 1964. In 1995 it hosted XV INCOSAI, the fifteenth triennial convention of the International Organization of Supreme Audit Institutions.[1] Funct...
Class of chemical compounds Diphosphate redirects here. It can also refer to any salt containing two phosphate groups. Pyrophosphate Phosphorus, P Oxygen, O Names IUPAC name Diphosphate Systematic IUPAC name Dipolyphosphate Other names PyrophosphatePhosphonatophosphate Identifiers CAS Number 14000-31-8 Y13472-36-1 (tetrasodium decahydrate salt) N7758-16-9 (disodium salt) N 3D model (JSmol) Interactive image ChEBI CHEBI:18361 Y ChemSpider 55...
Greek footballer (born 1993) Charalampos LykogiannisΧαράλαμπος Λυκογιάννης Lykogiannis with Greece U19 in 2012Personal informationFull name Charalampos Lykogiannis[1]Date of birth (1993-10-22) 22 October 1993 (age 30)[2]Place of birth Piraeus, GreeceHeight 1.90 m (6 ft 3 in)Position(s) Left-backTeam informationCurrent team BolognaNumber 22Youth career2007–2011 OlympiacosSenior career*Years Team Apps (Gls)2011–2015 Olympiacos 4 (0)20...
Slaven Bilić Bilic (kanan) dengan pelatih tim utama Edin Terzic Bulan Juli 2015Informasi pribadiTanggal lahir 11 September 1968 (umur 55)Tempat lahir Split, SFR YugoslaviaTinggi 1,89 m (6 ft 2+1⁄2 in)Posisi bermain BekInformasi klubKlub saat ini West Bromwich Albion (Pelatih)Karier junior1977–1988 Hajduk SplitKarier senior*Tahun Tim Tampil (Gol)1988–1993 Hajduk Split 109 (13)1988 → Primorac (pinjaman) 1988–1989 → Šibenik (pinjaman) 33 (7)1993–1996 Karl...
Tugu MudaLandskap daerah Tugu MudaInformasi umumLokasiSemarang, JatengAlamatJalan Pemuda Jalan Pandanaran Jalan MGR SoegijopranotoMulai dibangun10 November 1951Rampung20 Mei 1953Tinggi53 meterDesain dan konstruksiArsitekDr Mundo Hendro Tugu Muda (bahasa Jawa: ꦠꦸꦒꦸꦩꦸꦢ) adalah sebuah monumen yang dibuat untuk mengenang jasa-jasa para pahlawan yang telah gugur dalam Pertempuran Lima Hari di Semarang. Monumen ini terletak di Jalan Nasional Rute 20 yang mengarah ke Solo. Tugu ...
Puerto Rican rapper In this Spanish name, the first or paternal surname is Santiago and the second or maternal family name is Pérez. GuaynaaBackground informationBirth nameJean Carlos Santiago PérezBorn (1992-09-16) September 16, 1992 (age 31)Caguas, Puerto RicoGenres Reggaeton[1] LabelsUniversal LatinoRepublicSpouse(s)Lele Pons (m. 2023) [2]Musical artist Jean Carlos Santiago Pérez (born September 16, 1992), known professionally as Guaynaa, is a Puerto Rican r...
Військово-музичне управління Збройних сил України Тип військове формуванняЗасновано 1992Країна Україна Емблема управління Військово-музичне управління Збройних сил України — структурний підрозділ Генерального штабу Збройних сил України призначений для планува...
Peta infrastruktur dan tata guna lahan di Komune Royan. = Kawasan perkotaan = Lahan subur = Padang rumput = Lahan pertanaman campuran = Hutan = Vegetasi perdu = Lahan basah = Anak sungaiRoyanNegaraPrancis Royan adalah sebuah kota kecil dan merupakan salah satu commune dalam département Charente-Maritime, di sebelah barat Prancis. France. Kota ini terletak dekat kepulauan Ré, Oléron dan Île-d'Aix. Penduduknya pada tahun 1999 tercatat berjumla...
Overview of the impact of the COVID-19 pandemic on children This article needs to be updated. The reason given is: Have any long-term studies been conducted on this topic?. Please help update this article to reflect recent events or newly available information. (June 2023) Part of a series on theCOVID-19 pandemicScientifically accurate atomic model of the external structure of SARS-CoV-2. Each ball is an atom. COVID-19 (disease) SARS-CoV-2 (virus) Cases Deaths Timeline 2019 2020 January respo...
Irish footballer (born 1990) This article's lead section may be too short to adequately summarize the key points. Please consider expanding the lead to provide an accessible overview of all important aspects of the article. (May 2024) Paddy Madden Madden with Stockport County in 2023Personal informationFull name Patrick Stephen Madden[1]Date of birth (1990-03-04) 4 March 1990 (age 34)Place of birth Dublin, IrelandHeight 6 ft 0 in (1.83 m)Position(s) ForwardTeam inf...
Italian middle-distance runner Ala ZoghlamiAla Zoghlami at the 2020 Olympics.Personal informationNational team ItalyBorn (1994-06-19) 19 June 1994 (age 30)Tunis, TunisiaHeight1.80 m (5 ft 11 in)Weight57 kg (126 lb)SportSportAthleticsEvent(s)Middle-distance running3000 metres steeplechaseClubCus PalermoCoached byGaspare PolizziAchievements and titlesPersonal best 3000 m steeplechase 8:24.98 (2021) Ala Zoghlami (Arabic: علاء الزغلامي; born 19 June ...
Questa voce sull'argomento squadre di football americano norvegesi è solo un abbozzo. Contribuisci a migliorarla secondo le convenzioni di Wikipedia. NorvegiaSport Football americano FederazioneNoAFF ConfederazioneIFAF Europe Esordio internazionale Svezia 55 - 0 Norvegia Västerås, Svezia, 19 settembre 1997 Migliore vittoria Gran Bretagna 12 - 30 Norvegia Leeds, Regno Unito, 30 marzo 2019 Peggiore sconfitta Francia 84 - 0 Norvegia Nizza, Francia, 11 aprile 2004 Campionato europeo di f...
Katedral St. George'sKatedral Dikandung Tanpa Nodabahasa Inggris: Immaculate Conception CathedralKatedral St. George'sKoordinat: 12°3′6″N 61°45′19″W / 12.05167°N 61.75528°W / 12.05167; -61.75528LokasiSt. George'sNegaraGrenadaDenominasiGereja Katolik RomaArsitekturStatusKatedralStatus fungsionalAktifAdministrasiKeuskupanKeuskupan Saint George's Katedral St. George's yang bernama resmi Katedral Dikandung Tanpa Noda (bahasa Inggris: Immaculate Concepti...