Internal standard

In a chemical analysis, the internal standard method involves adding the same amount of a chemical substance to each sample and calibration solution. The internal standard responds proportionally to changes in the analyte and provides a similar, but not identical, measurement signal. It must also be absent from the sample matrix to ensure there is no other source of the internal standard present. Taking the ratio of analyte signal to internal standard signal and plotting it against the analyte concentrations in the calibration solutions will result in a calibration curve. The calibration curve can then be used to calculate the analyte concentration in an unknown sample.[1]

Selecting an appropriate internal standard accounts for random and systematic sources of uncertainty that arise during sample preparation or instrument fluctuation. This is because the ratio of analyte relative to the amount of internal standard is independent of these variations. If the measured value of the analyte is erroneously shifted above or below the actual value, the internal standard measurements should shift in the same direction.[1]

Ratio plot provides good way of compensation of detector sensitivity variation, but may be biased and should be replaced by Relative concentration/Relative calibration calculations if the reason of response variability is in different mass of analysed sample and traditional (not internal standard) calibration curve of any analyte is not linear through origin.[2]

History

The earliest recorded use of the internal standard method dates back to Gouy's flame spectroscopy work in 1877, where he used an internal standard to determine if the excitation in his flame was consistent.[3][4] His experimental procedure was later reintroduced in the 1940s, when recording flame photometers became readily available.[3] The use of internal standards continued to grow, being applied to a wide range of analytical techniques including nuclear magnetic resonance (NMR) spectroscopy, chromatography, and inductively coupled plasma spectroscopy.

Applications

Nuclear magnetic resonance spectroscopy

In NMR spectroscopy, e.g. of the nuclei 1H, 13C and 29Si, frequencies depend on the magnetic field, which is not the same across all experiments. Therefore, frequencies are reported as relative differences to tetramethylsilane (TMS), an internal standard that George Tiers proposed in 1958 and that the International Union of Pure and Applied Chemistry has since endorsed.[5][6] The relative difference to TMS is called chemical shift.[7]

TMS works as an ideal standard because it is relatively inert and its identical methyl protons produce a strong upfield signal, isolated from most other protons.[7] It is soluble in most organic solvents and is removable via distillation due to its low boiling point.[1]

In practice, the difference between the signals of common solvents and TMS are known. Therefore, no TMS needs to be added to commercial deuterated solvents, as modern instruments are capable of detecting the small quantities of protonated solvent present. By specifying the lock solvent to be used, modern spectrometers are able to correctly reference the sample; in effect, the solvent itself serves as the internal standard.[1]

Chromatography

In chromatography, internal standards are used to determine the concentration of other analytes by calculating response factor. The selected internal standard should have a similar retention time and derivatization. It must be stable and not interfere with the sample components. This mitigates the uncertainty that can occur in preparatory steps such as sample injection.[1]

In gas chromatography-mass spectrometry (GC-MS), deuterated compounds with similar structures to the analyte commonly act as effective internal standards.[8] However, there are non-deuterated internal standards such as norleucine, which is popular in the analysis of amino acids because it can be separated from accompanying peaks.[9][10][11]

Selecting an internal standard for liquid chromatography-mass spectrometry (LC-MS) depends on the employed ionization method. The internal standard needs a comparable ionization response and fragmentation pattern to the analyte.[12] LC-MS internal standards are often isotopically analogous to the structure of the analyte, using isotopes such as deuterium (2H), 13C, 15N and 18O.[13]

Inductively coupled plasma

Selecting an internal standard in inductively coupled plasma spectroscopy can be difficult, because signals from the sample matrix can overlap with those belonging to the analyte. Yttrium is a common internal standard that is naturally absent in most samples. It has both a mid-range mass and emission lines that don't interfere with many analytes. The intensity of the yttrium signal is what the signal from the analyte gets compared to.[1][14]

In Inductively coupled plasma-mass spectrometry (ICP-MS), species with a similar mass to the analyte usually serve as good internal standards, though not in every case. Factors that also contribute to the effectiveness of an internal standard in ICP-MS include how close its ionization potential, change in enthalpy, and change in entropy are to the analyte.[15]

Inductively coupled plasma-optical emission spectroscopy (ICP-OES) internal standards can be selected by observing how the analyte and internal standard signals change with varying experimental conditions. This includes making adjustments to the sample matrix or instrumentation settings and evaluating whether the selected internal standard is reacting in the same way the analyte is.[16]

Example of internal standard method

Spreadsheet for worked-example of plotting nickel concentrations in calibration solutions. The calibration curve on the top does not use the internal standard method. The calibration curve on the bottom does use the internal standard method.

One way to visualize the internal standard method is to create one calibration curve that doesn't use the method and one calibration curve that does. Suppose there are known concentrations of nickel in a set of calibration solutions: 0 ppm, 1.6 ppm, 3.2 ppm, 4.8 ppm, 6.4 ppm, and 8 ppm. Each solution also has 5 ppm yttrium to act as an internal standard. If these solutions are measured using ICP-OES, the intensity of the yttrium signal should be consistent across all solutions. If not, the intensity of the nickel signal is likely imprecise as well.

The calibration curve that does not use the internal standard method ignores the uncertainty between measurements. The coefficient of determination (R2) for this plot is 0.9985.

In the calibration curve that uses the internal standard, the y-axis is the ratio of the nickel signal to the yttrium signal. This ratio is unaffected by uncertainty in the nickel measurements, as it should affect the yttrium measurements in the same way. This results in a higher R2, 0.9993.

References

  1. ^ a b c d e f Skoog, Douglas A. (2018). Principles of instrumental analysis. F. James Holler, Stanley R. Crouch (7th ed.). Australia: Cengage Learning. ISBN 978-1-305-57721-3. OCLC 986919158.
  2. ^ Kalambet, Yuri; Kozmin, Yuri (18 December 2018). "Internal standard arithmetic implemented as relative concentration/relative calibration". Journal of Chemometrics. 33 (6): 1–7. doi:10.1002/cem.3106.
  3. ^ a b Burns, D. Thorburn; Walker, Michael J. (2019-05-01). "Origins of the method of standard additions and of the use of an internal standard in quantitative instrumental chemical analyses". Analytical and Bioanalytical Chemistry. 411 (13): 2749–2753. doi:10.1007/s00216-019-01754-w. ISSN 1618-2650. PMC 6522454. PMID 30941480.
  4. ^ Gouy, M. (1880). "Recherches photométriques sur les flammes colorée". Journal de Physique Théorique et Appliquée (in French). 9 (1): 19–27. doi:10.1051/jphystap:01880009001901. ISSN 0368-3893. S2CID 178492346.
  5. ^ Tiers, George Van Dyke (1958-09-01). "Reliable Proton Nuclear Resonance Shielding Values by "Internal Referencing" with Tetramethyl-silane". The Journal of Physical Chemistry. 62 (9): 1151–1152. doi:10.1021/j150567a041. ISSN 0022-3654.
  6. ^ Harris, Robin K.; Becker, Edwin D.; Cabral de Menezes, Sonia M.; Goodfellow, Robin; Granger, Pierre (2001-01-01). "NMR nomenclature. Nuclear spin properties and conventions for chemical shifts(IUPAC Recommendations 2001)". Pure and Applied Chemistry. 73 (11): 1795–1818. doi:10.1351/pac200173111795. ISSN 1365-3075. S2CID 95765050.
  7. ^ a b Guzman, Alexander L.; Hoye, Thomas R. (2022-01-21). "TMS is Superior to Residual C H Cl 3 for Use as the Internal Reference for Routine 1 H NMR Spectra Recorded in CDCl 3". The Journal of Organic Chemistry. 87 (2): 905–909. doi:10.1021/acs.joc.1c02590. ISSN 0022-3263. PMID 34974707. S2CID 245651135.
  8. ^ Unice, Kenneth M.; Kreider, Marisa L.; Panko, Julie M. (2012-11-08). "Use of a Deuterated Internal Standard with Pyrolysis-GC/MS Dimeric Marker Analysis to Quantify Tire Tread Particles in the Environment". International Journal of Environmental Research and Public Health. 9 (11): 4033–4055. doi:10.3390/ijerph9114033. ISSN 1660-4601. PMC 3524611. PMID 23202830.
  9. ^ Fico, D.; Margapoti, E.; Pennetta, A.; De Benedetto, G. E. (2018-04-01). "An Enhanced GC/MS Procedure for the Identification of Proteins in Paint Microsamples". Journal of Analytical Methods in Chemistry. 2018: e6032084. doi:10.1155/2018/6032084. ISSN 2090-8865. PMC 5902064. PMID 29805835.
  10. ^ Pons, Alexandre; Richet, Colette; Robbe, Catherine; Herrmann, Annkatrin; Timmerman, Philippe; Huet, Guillemette; Leroy, Yves; Carlstedt, Ingemar; Capon, Calliope; Zanetta, Jean-Pierre (2003-07-01). "Sequential GC/MS Analysis of Sialic Acids, Monosaccharides, and Amino Acids of Glycoproteins on a Single Sample as Heptafluorobutyrate Derivatives". Biochemistry. 42 (27): 8342–8353. doi:10.1021/bi034250e. ISSN 0006-2960. PMID 12846583.
  11. ^ Harduf, Z.; Bielorai, R.; Alumot, E. (1977-09-11). "Norleucine —an internal standard for the basic column used in physiological amino acid analysis". Journal of Chromatography A. 139 (1): 215–217. doi:10.1016/S0021-9673(01)84148-0. ISSN 0021-9673. PMID 893614.
  12. ^ "Internal Standard an Important Analyte Use in Drug Analysis by Liquid Chromatography Mass Spectrometry- An Article". International Journal of Pharmaceutical and Bio Medical Science. 2022-01-13. doi:10.47191/ijpbms/v2-i1-02. S2CID 245957199.
  13. ^ Grocholska, Paulina; Bąchor, Remigiusz (2021-05-18). "Trends in the Hydrogen−Deuterium Exchange at the Carbon Centers. Preparation of Internal Standards for Quantitative Analysis by LC-MS". Molecules. 26 (10): 2989. doi:10.3390/molecules26102989. ISSN 1420-3049. PMC 8157363. PMID 34069879.
  14. ^ Zachariadis, G. A.; Vogiatzis, C. (2010-05-17). "An Overview of the Use of Yttrium for Internal Standardization in Inductively Coupled Plasma–Atomic Emission Spectrometry". Applied Spectroscopy Reviews. 45 (3): 220–239. Bibcode:2010ApSRv..45..220Z. doi:10.1080/05704921003719122. ISSN 0570-4928. S2CID 98458774.
  15. ^ Finley-Jones, Haley J.; Molloy, John L.; Holcombe, James A. (2008-08-06). "Choosing internal standards based on a multivariate analysis approach with ICP(TOF)MS". Journal of Analytical Atomic Spectrometry. 23 (9): 1214–1222. doi:10.1039/B804048F. ISSN 1364-5544.
  16. ^ Sedcole, J. R.; Lee, J.; Pritchard, M. W. (1986-01-01). "Internal standard selection in the presence of matrix interactions in an inductively coupled argon plasma optimised for simultaneous multielement analysis by atomic emission spectrometry". Spectrochimica Acta Part B: Atomic Spectroscopy. 41 (3): 227–235. Bibcode:1986AcSpe..41..227S. doi:10.1016/0584-8547(86)80162-8. ISSN 0584-8547.

Read other articles:

Artikel ini tidak memiliki referensi atau sumber tepercaya sehingga isinya tidak bisa dipastikan. Tolong bantu perbaiki artikel ini dengan menambahkan referensi yang layak. Tulisan tanpa sumber dapat dipertanyakan dan dihapus sewaktu-waktu.Cari sumber: Tambak, Mojosongo, Boyolali – berita · surat kabar · buku · cendekiawan · JSTOR TambakDesaKantor Desa TambakNegara IndonesiaProvinsiJawa TengahKabupatenBoyolaliKecamatanMojosongoKode pos57322Kode Ke...

 

 

Kopal asal Madagaskar (Madagascar copalKopal merupakan getah (resin) aromatik yang dihasilkan oleh pohon kopal. Pohon ini merujuk pada beberapa spesies, seperti Protium copal, Hymenaea courbaril, Hymenaea verrucosa, Bursera coyucensis, Agathis dammara, dan sebagainya. Pohon kopal ditemukan di daerah tropis dan setiap spesies biasanya endemik di daerahnya. Hymenaea courbaril memiliki sebutan yang lebih khusus, yaitu pohon kopal Brazil,[1] sedangan Hymenaea verrucosa memiliki sebutan po...

 

 

Aichi Prefectural Ceramic Museum愛知県陶磁美術館Aichi Prefectural Ceramic Museum main buildingLocation within Aichi PrefectureShow map of Aichi PrefectureAichi Prefectural Ceramic Museum (Japan)Show map of JapanEstablished1 June 1978 (1978-06-01)LocationSeto, Aichi JapanCoordinates35°11′09″N 137°05′53″E / 35.18583°N 137.09806°E / 35.18583; 137.09806Collection size7020 itemsArchitectTaniguchi YoshirōOwnerAichi Prefecture Lobby of the ...

Surinamese football club Football clubRandjiet BoysFull nameSociaal Cultureel en Sportvereniging Randjiet BoysGroundDr. Ir. Franklin Essed Stadium, Paramaribo, SurinameCapacity35002013–14Eerste Klasse, 11th (relegated) Home colours SCS Randjiet Boys is a Surinamese football club. They play their home games at the 3,500-capacity Dr. Ir. Franklin Essed Stadium in Paramaribo.[1] They were relegated from the Hoofdklasse in 2012–13[2] and again relegated, from the Eerste Klasse...

 

 

Синелобый амазон Научная классификация Домен:ЭукариотыЦарство:ЖивотныеПодцарство:ЭуметазоиБез ранга:Двусторонне-симметричныеБез ранга:ВторичноротыеТип:ХордовыеПодтип:ПозвоночныеИнфратип:ЧелюстноротыеНадкласс:ЧетвероногиеКлада:АмниотыКлада:ЗавропсидыКласс:Пт�...

 

 

This article relies excessively on references to primary sources. Please improve this article by adding secondary or tertiary sources. Find sources: List of Saturday Night Live commercial parodies – news · newspapers · books · scholar · JSTOR (January 2023) (Learn how and when to remove this template message) On the American late-night live television sketch comedy and variety show Saturday Night Live (SNL), a commercial advertisement parody is common...

MućcomuneLocalizzazioneStato Croazia Regione Spalatino-dalmata TerritorioCoordinate43°43′34″N 16°30′00″E / 43.726111°N 16.5°E43.726111; 16.5 (Muć)Coordinate: 43°43′34″N 16°30′00″E / 43.726111°N 16.5°E43.726111; 16.5 (Muć) Superficie210,80 km² Abitanti3 882[2] (2011) Densità18,42 ab./km² Altre informazioniCod. postale21255 Prefisso021 Fuso orarioUTC+1 CartografiaMuć Sito istituzionaleModifica dati su...

 

 

National emblem of Maldives Emblem of MaldivesArmigerRepublic of MaldivesShieldA star and crescent or atop a coconut palm properSupportersThe National Flag of Maldives on either sideMottoالدولة المحلديبية State of the Mahal Dibiyat The Maldivian National Emblem[1] consists of a coconut palm, a crescent, and two criss-crossing National Flags with the traditional Title of the State. Interpretation The depicted coconut palm represents the livelihood of the Nation accordin...

 

 

2016年美國總統選舉 ← 2012 2016年11月8日 2020 → 538個選舉人團席位獲勝需270票民意調查投票率55.7%[1][2] ▲ 0.8 %   获提名人 唐納·川普 希拉莉·克林頓 政党 共和黨 民主党 家鄉州 紐約州 紐約州 竞选搭档 迈克·彭斯 蒂姆·凱恩 选举人票 304[3][4][註 1] 227[5] 胜出州/省 30 + 緬-2 20 + DC 民選得票 62,984,828[6] 65,853,514[6]...

国民阵线Barisan NasionalNational Frontباريسن ناسيونلபாரிசான் நேசனல்国民阵线标志简称国阵,BN主席阿末扎希总秘书赞比里署理主席莫哈末哈山总财政希山慕丁副主席魏家祥维纳斯瓦兰佐瑟古律创始人阿都拉萨成立1973年1月1日 (1973-01-01)[1]设立1974年7月1日 (1974-07-01)前身 联盟总部 马来西亚  吉隆坡 50480 秋傑区敦依斯迈路太子世贸中心(英�...

 

 

King of the Foreigners Amlaíb ConungKing of the ForeignersReignc. 853–871Diedc. 874IssueOistinCarlusFatherGofraid Amlaíb Conung (Old Norse: Óláfr [ˈoːˌlɑːvz̠]; died c. 874) was a Viking[nb 1] leader in Ireland and Scotland in the mid-late ninth century. He was the son of the king of Lochlann, identified in the non-contemporary Fragmentary Annals of Ireland as Gofraid, and brother of Auisle and Ímar, the latter of whom founded the Uí Ímair dynasty, and whose...

 

 

Railway station in Amsterdam, Netherlands Amsterdam ArenAGeneral informationLocationNetherlandsCoordinates52°18′58″N 4°56′34″E / 52.31611°N 4.94278°E / 52.31611; 4.94278Operated byNederlandse SpoorwegenPlatforms1 (train)Other informationStation codeAsdarHistoryOpened28 November 1996LocationAmsterdam ArenALocation within Northern RandstadShow map of Northern RandstadAmsterdam ArenAAmsterdam ArenA (Netherlands)Show map of Netherlands Halte Amsterdam ArenA (As...

This article includes a list of references, related reading, or external links, but its sources remain unclear because it lacks inline citations. Please help improve this article by introducing more precise citations. (July 2014) (Learn how and when to remove this message) Academic journalThe Contemporary PacificDisciplinePacific studiesLanguageEnglishPublication detailsHistory1989–presentPublisherUniversity of Hawaii Press (USA)FrequencysemiannualStandard abbreviationsISO 4 (alt) ...

 

 

British Army operation in Belfast in 1970 Falls CurfewPart of the Troubles/Operation BannerA mural depicting the march that broke through the curfewDate3–5 July 1970LocationBelfast, Northern IrelandResult British victoryCurfew broken in the aftermath Large number of IRA weapons captured 337 people arrested Curfew broken by neighbouring residents 4 civilians killed & several other civilians injured by the British ArmyBelligerents British Government Official IRA Provisional IRACommanders ...

 

 

Food and agriculture in Nazi Germany describes the food and agricultural policies of Nazi Germany and their consequences from 1933 when the Nazis took power in Germany until 1945 when Germany was defeated in World War II (1939–1945) by the allied nations. Starvation and its associated illnesses killed about 20 million people in Europe and Asia during World War II, approximately the same as the number of soldiers killed in battle.[1] Most of the deaths from starvation in Europe were ...

Cricket team Chilaw Marians Cricket ClubPersonnelCaptain Pulina TharangaCoach Sarath JayawardenaTeam informationColoursCream/pink  [1]Founded1975Home groundFTZ Sports ComplexCapacity1,000HistoryFirst-class debutin 2000Premier League Tournament wins2Premier Limited Overs Tournament wins0Twenty20 Tournament wins1 (2004)Official websitewww.chilawmarians.com Chilaw Marians Cricket Club are a first-class cricket team based in Chilaw, Sri Lanka. They play their home...

 

 

Ballet ManilaGeneral informationNameBallet ManilaYear founded1995Principal venuePasay, PhilippinesWebsiteballetmanila.com.phArtistic staffArtistic DirectorLisa Macuja-Elizalde (Artistic Director and Principal Dancer)Resident ChoreographersGerardo Francisco (Resident Choreographer and Rehearsal Master for Modern Dance)Other Artistic StaffOsias Barroso (Artistic Associate & Rehearsal Master)Jonathan Janolo (Assistant Ballet Master/Set & Costume Designer)Eileen Lopez (Assistant Rehearsal...

 

 

Heinrich Wilhelm Matthäus Olbers Nama dalam bahasa asli(de) Heinrich Wilhelm Matthias Olbers BiografiKelahiran11 Oktober 1758 Hemelingen (Kekaisaran Romawi Suci) Kematian2 Maret 1840 (81 tahun)Bremen (Konfederasi Jerman) Tempat pemakamanBremen Galat: Kedua parameter tahun harus terisi! Member of the French National Assembly (en) Data pribadiPendidikanUniversitas Georg August Göttingen - kedokteran (1777–)Athenaeum (en) KegiatanSpesialisasiAstronomi Pekerjaandokter (1781–182...

此條目需要补充更多来源。 (2024年6月4日)请协助補充多方面可靠来源以改善这篇条目,无法查证的内容可能會因為异议提出而被移除。致使用者:请搜索一下条目的标题(来源搜索:金源镇 — 网页、新闻、书籍、学术、图像),以检查网络上是否存在该主题的更多可靠来源(判定指引)。 金源镇(朝鮮語:김원진 Kim Wonjin,1992年5月1日—),韩国男子柔道运动员,201...

 

 

Time zone SAST redirects here. For source code review, see Static application security testing. For the Chinese space agency, see Shanghai Academy of Spaceflight Technology. Time zones of Africa: Light Blue Cape Verde Time[a] (UTC−1) Blue Greenwich Mean Time (UTC) Red Central European TimeWest Africa Time (UTC+1) Ochre Central Africa TimeEastern European TimeEgypt Standard TimeSouth African Standard Time (UTC+2) Green East Africa Time (UTC+3) Turquoise Mauritius Time[b]Seychelles Time[b] (U...