Endodontic files and reamers

Endodontic files and reamers are surgical instruments used by dentists when performing root canal treatment. These tools are used to clean and shape the root canal, with the concept being to perform complete chemomechanical debridement of the root canal to the length of the apical foramen. Preparing the canal in this way facilitates the chemical disinfection to a satisfactory length but also provides a shape conducive to obturation (filling of the canal).

Hand files

Miscellaneous endodontic instruments. From left: Lentulo spiral, reamer, K-file and H-file.

Hand files can provide tactile sensation when cleaning or shaping root canals. This allows the dentist to feel changes in resistance or angulation, which can help determine curvature, calcification and/or changes in anatomy, in which two dimensional radiographs may not always identify. This information can help determine strategies or avoid complications before moving on to rotary instruments.

K-type files

The cutting edge of K type files is made up of twisted squares of stainless steel alloy. The K-flex file differs for the fact it has a rhomboid shaped cross-section and has an increased flexibility compared to traditional K-files.[1]

C-type files

C-files are stiffer than K-files, and are recommended for calcified canals and ones that are curved and narrow.[2]

Nickel-titanium files

Nickel-titanium is a superelastic alloy which allows it to undergo greater stresses compared to stainless steel therefore files have a reduced risk of file fracture. It also has the characteristic of 'shape memory' which allows it to return to its initial shape through heating after strain. This reduces the risk of deformation within the root canal as forces of compression and tension are absent.

The superelasticity allows an increase in taper (between 4–8%) compared to stainless steel. This allows an adequate taper of the root canal which takes less time to prepare than with stainless steel and less files needed. The super elasticity also means the risk of zipping and apical transportation is reduced.

Many Nickel-titanium files are available. The files can be used within rotary systems or manually for a higher level of control.

Techniques for use

Watch winding and circumferential filing technique

The use of the file in a forwards and backwards motion, as if watch winding, with slight apical pressure. This allows the file to effectively debride the canal dentine by moving slowly down the canal.

For K-type files, once the file has reached the desired working length, a push and pulling action is used around the circumference of the canal, while only maintaining contact with the canal wall on the outstroke to minimise a debris blockage apically.

The balanced force technique

This is the most widely used technique and especially good for working with curved canals.[3]

Files used for this technique need to be non-cutting edge and flexible. The file is rotated 60 degrees clockwise in the canal when a slight resistance is felt. The file is then rotated 360 degrees anticlockwise to pick up the dentine in the flutes that was made during the first rotation. This should be done no more than three times before the file is removed and cleaned and the canal system irrigated before reinsertion.[4]

Hedstrom files

The cross-section of a Hedstrom file (H-file) is made up of a continuous sequence of cones. They are very sharp with a cutting tip. Their use in a push-pull fashion results in a high level of debridement on removal from the root canal. They should not be rotated more than 30 degrees as they are narrow and vulnerable to fracture. They are also used for removal of root canal filling materials e.g. gutta percha during secondary root canal treatment.

Barbed broach

This file is used to remove pulp tissue (extirpation) during root canal treatment. There are sharp barbs on the file to engage the pulp tissue and remove this efficiently. These files are not used to shape the RCS.

Standardisation of instruments (ISO)

The handles of the ISO instruments are colour coded and are available in three different lengths of 21mm, 25mm and 31mm where the extra length is non-cutting shaft. This extra length is particularly useful for posterior teeth where access and visibility is impaired.

ISO files are made of stainless steel. This can be useful in smaller files (<20) but larger files have increased rigidity which can result in procedural errors. At smaller sizes the files can be pre-curved which is a major advantage for the debridement of roots with sharp curvatures. Their rigidity also has an advantage in calcified root canals in the initial stages of debridement. 

The ISO stainless steel files on the market today include K-Flex, K-Flexofile and Hedström where the tip size and taper is standardised.

ISO normed hand files have a standardised taper of 2% that equates to 0.02mm increase in diameter per mm of file. This standardised taper allows you to calculate the diameter of any given stainless steel file at any given point. Where the 2% taper means that there is an increase in diameter by 0.02mm every 1mm of file (moved in a coronal direction). The most apical point of any file is deemed D0, so moving coronal on the file by 1mm brings you to D1 and so on, up to D16 as there is a 16mm cutting surface on all files.

For example, an ISO K file size 25 has a D0 value of 0.25mm diameter at its tip. If you were to move 6mm coronally on this file from D0, the cross sectional diameter would be:

0.25mm + (6mmx0.02mm)=0.37mm

Protaper series

The range of files are available as hand and rotary. The first files in the series are termed SX, S1 and S2. These are used to improve access to the canals by first creating a coronal flare in the crown-down technique.

  • SX files: D0 value of 0.19mm
  • S1 files: D0 value of 0.17mm
  • S2 files: D0 value of 0.20mm

SX files are typically used first as they are shorter in overall length 19mm and so are good in cases of restrictive space. The canal is prepped in the coronal 2/3 with these files as part of the crown-down technique.

After this, files named F1, F2, F3 etc. are used with increasing D0 values. These are used to shape the canal.

  • F1 files: D0 value of 0. 20mm
  • F2 files: D0 value of 0.25mm
  • F3 files: D0 value of 0.30mm etc.

Between each of these finishing files, you should recapitulate the canal using the corresponding (with the same D0 value) K file. This prevents procedural errors, confirms the canal remains patent and prevents dentine swarf build up in within the canal. Complete copious irrigation in between each file.

Rotary files

Revo-S file system SC1, SC2 and SU shaping files by Micromega

The introduction of Nickel Titanium in dentistry has allowed the use of rotary systems to be used to prepare root canals safely and predictably. Rotary instrumentation is known to have an improved cutting efficiency when compared with hand filing techniques. It is advisable to use a dedicated electric endodontic motor where torque and speed can be easily controlled dependent on the system chosen. Despite the advantages of rotary systems, it is always recommended to create a glide path with hand files in each canal prior to rotary instrumentation. There are numerous rotary files available on the market, including a variety of systems from different manufacturers.

Reciprocating systems

Reciprocating systems involve rotation of the file in both anti-clockwise and clockwise directions. This is similar to the ‘balanced force’ mechanism used with hand files. When the file is used in an anti-clockwise direction, it engages dentine and is quickly followed by a clockwise turn before re-engaging the root canal wall and shearing the dentine. Benefits of a reciprocating system include:

  • Reduced risk of cyclical failure
  • Reduced risk of torsional failure
  • Simple protocol with single file (small, regular or large based on canal size) therefore more cost effective

Self-adjusting files

Self-adjusting file systems have been developed to overcome complications that arise due to complex anatomy and canal configurations. These files are used in a rotary hand piece and consist of a flexible, thin NiTi lattice with a hollow centre that adapt three-dimensionally to the shape of a given root canal, including its cross section.[5] The files are operated with vibratory in-and-out motion, with continuous irrigation of disinfectant delivered by a peristaltic pump through the hollow file.[6] A uniform layer of dentin is removed from the whole circumference of the root canal, thus achieving the main goals of root canal treatment while preserving the remaining root dentin.[7] The 3D scrubbing effect of the file, combined with the fresh irrigant, result in clean canals, which in turn facilitate better obturation.[5] More effective disinfection of flat-oval root canals is another goal that is simultaneously attained.[8]

D-files

D files are a selection of bespoke rotary files that are commonly used in re-treatment cases for the efficient removal of gutta percha. They are used in sequence to remove the coronal (D1), mid (D2) and apical (D3) ⅓ root filling material more efficiently before the final shaping with conventional instruments. D1 is 16mm in length with a cutting end tip to engage the filling material in the canal. D2 and D3 are 18mm and 22mm in length respectively, both are non end cutting and aim to not remove remaining dentine from canal walls in the process.[1]

Single use legislation (in the UK)

In 2007, new legislation documenting the possible risk of prion disease transmission via endodontic files/reamers during root canal treatment was published via the BDJ.[9] The conclusions made were such that there was no significant risk associated but the implementation of single use instruments was introduced to take all possible precautions. This was primarily due to the shape and relative surface area of the files making thorough disinfection and sterilisation very difficult.

Mechanisms of failure

Instrumentation of the root canal systems (RCS) can lead to procedural errors including ledging, zipping, canal perforation and apex transportation all of which can be somewhat successfully resolved through further manual corrective techniques. However, file separation whereby the instrument breaks in the canal, is the most concerning and problematic procedural error, with fractured endodontic instruments being the most commonly found object in the RCS. The incidence of file fracture has been found to range between 0.25-6% of cases. File separation will create an obstruction within the canal preventing adequate cleaning and shaping of the canal at and beyond the obstruction as well as under-filling of the RCS. This may ultimately lead to endodontic failure depending on the location at which the file fractured in the RCS.

The cause fracture of instruments can be divided into different factors, operator/ technique, anatomy and instrument.

Cyclic fatigue

i.e. the lack of flexibility of the instruments when negotiating particularly curved canals. The more curved the canal, the greater the cyclical fatigue placed on the instrument, as it is undergoing repetitive tensile and compressive stresses upon rotation no matter the flexibility of the alloy. Pre-curving of the stainless steel files for canal negotiation will work-harden them, rendering them more brittle and therefore are more likely to fracture. Such files should also not be twisted in an anticlockwise manner, as this may also lead to brittle fracture especially when there is increased torque. NiTi files have been designed with increased flexibility for canal negotiation, however this does not entirely negate the event of file separation. NiTi files undergo cyclic fatigue due to a change in the crystalline structure of the file whilst under stress resulting in the alloy becoming more brittle.

Flexural fatigue

i.e. overuse of the file. It is safe to assume that the more a file is used, the greater the risk of separation. However, one cannot dictate a specific number of times for use nor predict when a file is going to fracture. The introduction of single use files has reduced this risk somewhat, yet it is vital to regularly inspect the files upon removal from canals for damage. The problem comes when files separate without there being any visible sign of damage.

Torsional fatigue

Torque relates to the required force needed in order for an instrument to carry on rotating upon encountering frictional forces. A file may bind the wall of the root canal apically due to a larger diameter of the file compared with the canal causing friction. If rotational forces are still in motion, torque may reach a critical level and the file will fracture. The torque generated from smaller canals will be greater than in larger canals, as files will bind to the canal walls more readily through friction. The greater the diameter of the instrument, the more force it can withstand despite needing increased torque however, the less resistant it becomes to cyclic fatigue. Torsional fatigue can be somewhat limited through creation of a glide path and adopting the Crown-Down technique in a bid to reduce frictional forces.

Intrinsic file defects

Beware of surface defects arising from the manufacture of the files, which can propagate under fatigue by creating stress concentrations and ultimately lead to fracture. This holds especially true for NiTi files, which are manufactured via milling of alloy blanks using CAD-CAM, as opposed to twisting of the blanks like with stainless steel. Deeper cutting flutes will also create stress concentrations.

File failure could be attributed to the skill and chosen technique used for instrumentation by the operator. It is more often the way in which an instrument is used, as opposed to the number of times it has been used that causes fracture e.g. due to overloading. Aggressively inserting instruments into canals should be avoided, as this will increase the friction created between the canal walls and the file. Evidence shows that hand instrumentation will result in a lower risk of file fracture compared with rotary and this may be attributed to increased rotational speed, which enhances the effects of cyclic fatigue. Therefore, when using electric motors with rotary instruments, a low speed and low torque concept is recommended.

Minimising the risk of separation

  1. Well-angled radiographs to determine canal curvature (this will however be a 2D representation of a 3D system)
  2. Access cavity design (straight line access) and glide path
  3. Crown Down instrumentation sequence to minimise friction
  4. Wet canals for lubrication but beware of risk of corrosion to stainless steel instruments due to irrigants used in canals e.g. with EDTA or Sodium Hypochlorite
  5. Regular file inspection before and during instrumentation
  6. Set electric motors at low torque (follow manufacturer instruction for recommended speed and torque)

References

  1. ^ a b Patel, S. Barnes, J (2013). The principles of Endodontics. 2nd ed. Oxford: Oxford University Press. 69-72.
  2. ^ Peter Mann DDS, FOCOI, FAGD https://www.perioimplantadvisory.com/articles/2015/10/step-by-step-procedure-to-simplified-and-efficient-root-canal-techniques.html
  3. ^ "Endodontics: Part 7 Preparing the root canal". British Dental Journal. Retrieved 2017-11-21.
  4. ^ "Endodontics: Part 7 Preparing the root canal". British Dental Journal. Retrieved 2017-11-21.
  5. ^ a b Metzger, Zvi; Teperovich, Ehud; Zary, Raviv; Cohen, Raphaela; Hof, Rafael (2010). "The Self-adjusting File (SAF). Part 1: Respecting the Root Canal Anatomy—A New Concept of Endodontic Files and Its Implementation". Journal of Endodontics. 36 (4): 679–90. doi:10.1016/j.joen.2009.12.036. PMID 20307744.
  6. ^ Metzger, Zvi (2014). "The self-adjusting file (SAF) system: An evidence-based update". Journal of Conservative Dentistry. 17 (5): 401–419. doi:10.4103/0972-0707.139820. ISSN 0972-0707. PMC 4174698. PMID 25298639.
  7. ^ De-Deus, Gustavo; Souza, Erick Miranda; Barino, Bianca; Maia, Janaina; Zamolyi, Renata Quintella; Reis, Claudia; Kfir, Anda (2011). "The Self-Adjusting File Optimizes Debridement Quality in Oval-shaped Root Canals". Journal of Endodontics. 37 (5): 701–5. doi:10.1016/j.joen.2011.02.001. PMID 21496675.
  8. ^ Siqueira Jr., José F.; Alves, Flávio R. F.; Almeida, Bernardo M.; Machado De Oliveira, Julio C.; Rôças, Isabela N. (2010). "Ability of Chemomechanical Preparation with Either Rotary Instruments or Self-adjusting File to Disinfect Oval-shaped Root Canals". Journal of Endodontics. 36 (11): 1860–5. doi:10.1016/j.joen.2010.08.001. PMID 20951301.
  9. ^ "Government advises single use of endodontic instruments" (PDF). British Dental Journal. 202 (8): 442. 2007. doi:10.1038/bdj.2007.364.

Read other articles:

Artikel ini sebatang kara, artinya tidak ada artikel lain yang memiliki pranala balik ke halaman ini.Bantulah menambah pranala ke artikel ini dari artikel yang berhubungan atau coba peralatan pencari pranala.Tag ini diberikan pada November 2022. Jean-Daniel SimonLahir(1942-11-30)30 November 1942Salon-de-Provence, Bouches-du-Rhône, PrancisMeninggal3 Februari 2021 (usia 78)Paris, PrancisPekerjaanSutradara, penulis naskahTahun aktif1968–1985 Jean-Daniel Simon (30 November 1942 ̵...

 

City in Texas, United StatesWhite Settlement, TexasCityCity of White SettlementLocation of White Settlement in Tarrant County, TexasCoordinates: 32°45′33″N 97°27′38″W / 32.75917°N 97.46056°W / 32.75917; -97.46056Country United StatesState TexasCounty TarrantGovernment • TypeCouncil-Manager • City CouncilMayor Faron Young Amber Munoz (Mayor pro tempore) Paul Moore Evelyn Spurlock Amber Munoz Gregg Geesa William Wright  ...

 

Turbo Pascal Layar pembuka Turbo Pascal 4.0 (1987).Tipebahasa pemrograman, Lingkungan pengembangan terpadu dan bahasa pemrograman berorientasi obyek Versi pertama20 November 1983 GenreIntegrated Development EnvironmentKarakteristik teknisSistem operasiDOS, CP/M, CP/M-86 (en), Windows 3.x dan Classic Mac OS Bahasa pemrogramanBahasa rakitan Format berkasDaftarTurbo Pascal Desktop (en), Turbo Pascal configuration (en), Borland Turbo Pascal 5.5 compiled Unit (en), Borland Turbo Pascal 6.0 compile...

artikel ini tidak memiliki pranala ke artikel lain. Tidak ada alasan yang diberikan. Bantu kami untuk mengembangkannya dengan memberikan pranala ke artikel lain secukupnya. (Pelajari cara dan kapan saatnya untuk menghapus pesan templat ini) Artikel ini sebatang kara, artinya tidak ada artikel lain yang memiliki pranala balik ke halaman ini.Bantulah menambah pranala ke artikel ini dari artikel yang berhubungan atau coba peralatan pencari pranala.Tag ini diberikan pada Maret 2016. Pengukuran Ke...

 

كلية الخليل للتمريض معلومات التأسيس 1994 الموقع الجغرافي المدينة الخليل ، فلسطين المكان الضفة الغربية، محافظة الخليل البلد  فلسطين إحصاءات تعديل مصدري - تعديل   كلية الخليل للتمريض والتي تمثل إحدى الكليات الرائدة في مجال التمريض بفلسطين والتي تسعى جاهده بخطوات ايجابي...

 

Sporting event delegationIraq at the2020 Summer ParalympicsFlag of IraqIPC codeIRQNPCIraqi National Paralympic Committeein Tokyo, JapanAugust 24, 2021 (2021-08-24) – September 5, 2021 (2021-09-05)Competitors19 in 5 sportsMedals Gold 0 Silver 1 Bronze 2 Total 3 Summer Paralympics appearances (overview)199219962000200420082012201620202024 Iraq competed at the 2020 Summer Paralympics in Tokyo, Japan, from 24 August to 5 September 2021.[1] Medalists Me...

Industries involved in the production and sale of energy The energy industry is the totality of all of the industries involved in the production and sale of energy, including fuel extraction, manufacturing, refining and distribution. Modern society consumes large amounts of fuel, and the energy industry is a crucial part of the infrastructure and maintenance of society in almost all countries. In particular, the energy industry comprises: the fossil fuel industries, which include petroleum i...

 

Diffuse luminance of the night sky This article is about night sky luminance from artificial light. For the natural atmospheric phenomenon, see airglow. For the interplanetary phenomenon, see zodiacal light. For the wider environmental impacts of artificial light, see light pollution. Mexico City at night, showing skyglow A map from 1996 to 1997 showing the extent of skyglow over Europe Skyglow (or sky glow) is the diffuse luminance of the night sky, apart from discrete light sources such as ...

 

Частина серії проФілософіяLeft to right: Plato, Kant, Nietzsche, Buddha, Confucius, AverroesПлатонКантНіцшеБуддаКонфуційАверроес Філософи Епістемологи Естетики Етики Логіки Метафізики Соціально-політичні філософи Традиції Аналітична Арістотелівська Африканська Близькосхідна іранська Буддій�...

Частина серії проФілософіяLeft to right: Plato, Kant, Nietzsche, Buddha, Confucius, AverroesПлатонКантНіцшеБуддаКонфуційАверроес Філософи Епістемологи Естетики Етики Логіки Метафізики Соціально-політичні філософи Традиції Аналітична Арістотелівська Африканська Близькосхідна іранська Буддій�...

 

Carex phaeocephala Klasifikasi ilmiah Kerajaan: Plantae (tanpa takson): Tracheophyta (tanpa takson): Angiospermae (tanpa takson): Monokotil (tanpa takson): Komelinid Ordo: Poales Famili: Cyperaceae Genus: Carex Spesies: Carex phaeocephala Nama binomial Carex phaeocephalaPiper Carex phaeocephala adalah spesies tumbuhan seperti rumput yang tergolong ke dalam famili Cyperaceae. Spesies ini juga merupakan bagian dari ordo Poales. Spesies Carex phaeocephala sendiri merupakan bagian dari genus Car...

 

Целые числа от нуля до ста. Простые числа отмечены красным. Разложение числа 42 на простые множители: 42 = 2 × 3 × 7 {\displaystyle 42=2\times 3\times 7} Просто́е число́ — натуральное число, имеющее ровно два различных натуральных делителя. Другими словами, натуральное число p {\displaystyle...

British bimonthly journal Academic journalNew Left ReviewDisciplinePoliticsLanguageEnglishEdited bySusan WatkinsPublication detailsHistory1960–presentPublisherNew Left Review Ltd (United Kingdom)FrequencyBimonthlyImpact factor1.967 (2018)Standard abbreviationsISO 4 (alt) · Bluebook (alt1 · alt2)NLM (alt) · MathSciNet (alt )ISO 4New Left Rev.IndexingCODEN (alt · alt2) · JSTOR (alt) · LCCN (alt)MIAR&...

 

Untuk kepala pemerintahan Hindia Belanda, lihat Wali Kota Batavia. Gubernur Daerah Khusus Ibukota JakartaPetahanaHeru Budi HartonoPenjabatsejak 17 Oktober 2022Pemerintah Provinsi Daerah Khusus Ibukota JakartaKediamanRumah Dinas Gubernur Daerah Khusus Ibukota Jakarta, Menteng, Jakarta PusatMasa jabatan5 tahun; dapat diperpanjang sekaliPejabat perdanaSuwiryo (Wali Kota Jakarta Raya)Dibentuk7 September 1945; 78 tahun lalu (1945-09-07)WakilWakil Gubernur Daerah Khusus Ibukota JakartaSit...

 

2006 DP14PenemuanDitemukan olehLINEARSitus penemuan704Tanggal penemuan2006/02/23Ciri-ciri orbitAphelion2.425Perihelion0.306Sumbu semimayor1.365Eksentrisitas0.776Anomali rata-rata326.3Inklinasi11.8Bujur node menaik317.3Argumen perihelion59.1Ciri-ciri fisikMagnitudo mutlak (H)18.8 2006 DP14 adalah sebuah asteroid. Asteroid ini merupakan bagian dari asteroid Apollo, yang terletak dekat dengan bumi. Eksentrisitas orbit asteroid ini tercatat sebesar 0.776, sementara magnit...

Questa voce o sezione sull'argomento modelli non cita le fonti necessarie o quelle presenti sono insufficienti. Puoi migliorare questa voce aggiungendo citazioni da fonti attendibili secondo le linee guida sull'uso delle fonti. Segui i suggerimenti del progetto di riferimento. Freja Beha ErichsenAltezza178[1] cm Misure82-63-86[1] Taglia34[1] (UE) Peso59 kg Scarpe39[1] (UE) OcchiCastani CapelliCastano chiaro Modifica dati su Wikidata · Manu...

 

Al-MurjanLingkunganNegaraArab SaudiProvinsiProvinsi MakkahPemerintahan • Wali kotaHani Abu Ras[1] • Gubernur kotaMish'al Al-SaudKetinggian12 m (39 ft)Zona waktuUTC+3 (AST) • Musim panas (DST)ASTKode pos(5 kode digit dimulai dari 23; e.g. 23434)Kode area telepon+966-12Situs webwww.jeddah.gov.sa/english/index.php Al-Murjan adalah sebuah pemukiman padat penduduk di kota Jeddah di Provinsi Makkah, tepatnya di sebelah barat Arab Saudi.[3 ...

 

Cold-growing Cymbidium Klasifikasi ilmiah Kerajaan: Plantae (tanpa takson): Tracheophyta (tanpa takson): Angiospermae (tanpa takson): Monokotil Ordo: Asparagales Famili: Orchidaceae Subfamili: Epidendroideae Tribus: Cymbidieae Subtribus: Cyrtopodiinae Alliance: Cymbidium Genus: Cymbidium Spesies: C. kanran Nama binomial Cymbidium kanranMakino (1902) Sinonim Cymbidium kanran f. purpurascens Makino (1902) Cymbidium oreophilum Hayata (1914) Cymbidium purpureohiemale Hayata (1914) Cymbidium...

For the Game of Thrones episode, see A Golden Crown. Imperial Iranian Air Force Aerobatic Team Golden CrownGolden Crown insigniaActive30 October 1958-11 February 1979Disbandedafter 1979 Revolution (revival is still ongoing)Country IranBranchImperial Iranian Air ForceRoleAerobatic display teamPart ofpart of 1st Fighter Interceptor SquadronGarrison/HQMehrabad (THR)Khatami Air BaseColours■□■(three colours of the flag of Iran)CommandersLeaderNader Jahanbani (1958-1962,1969)Amir Hossein...

 

Split Screen einer Videoüberwachungsanlage Split Screen oder Bildschirmaufteilung (wörtlich „geteilter Bildschirm“) ist eine in visuellen Medien verwendete Technik, die das Bild auf dem Bildschirm (engl. screen) in zwei (oder mehr) Bereiche aufteilt (engl. splittet), um zwei oder mehr Geschehnisse oder Bilder gleichzeitig zu zeigen. Inhaltsverzeichnis 1 Film 2 Duo-Vision 3 Nachrichten 4 Sportübertragungen 5 Computerspiele 6 Anzeigegeräte 7 Siehe auch 8 Literatur 9 Einzelnachweise Film...