Blood pressure measurement

Blood pressure measurement
A medical student checking blood pressure using a sphygmomanometer and stethoscope.
Test ofBlood pressure
Based onCNAP vascular unloading technique
MedlinePlus007490
Right position for taking blood pressure
Digital blood pressure monitor in use

Arterial blood pressure is most commonly measured via a sphygmomanometer, which historically used the height of a column of mercury to reflect the circulating pressure.[1] Blood pressure values are generally reported in millimetres of mercury (mmHg), though modern aneroid and electronic devices do not contain mercury.

For each heartbeat, blood pressure varies between systolic and diastolic pressures. Systolic pressure is peak pressure in the arteries, which occurs near the end of the cardiac cycle when the ventricles are contracting. Diastolic pressure is minimum pressure in the arteries, which occurs near the beginning of the cardiac cycle when the ventricles are filled with blood. An example of normal measured values for a resting, healthy adult human is 120 mmHg systolic and 80 mmHg diastolic (written as 120/80 mmHg, and spoken as "one-twenty over eighty"). The difference between the systolic and diastolic pressures is referred to as pulse pressure (not to be confused with pulse rate/heartrate) and has clinical significance in a wide variety of situations. It is generally measured by first determining the systolic and diastolic pressures and then subtracting the diastolic from the systolic.[2][3][4][5] Mean arterial pressure is the average pressure during a single cardiac cycle and, although it is possible to measure directly using an arterial catheter, it is more commonly estimated indirectly using one of several different mathematical formulas once systolic, diastolic, and pulse pressures are known.[6]

Systolic and diastolic arterial blood pressures are not static but undergo natural variations[7] from one heartbeat to another and throughout the day (in a circadian rhythm). They also change in response to stress, nutritional factors, drugs, disease, exercise, and momentarily from standing up. Sometimes the variations are large. Hypertension refers to arterial pressure being abnormally high, as opposed to hypotension, when it is abnormally low. Along with body temperature, respiratory rate, and pulse rate, blood pressure is one of the four main vital signs routinely monitored by medical professionals and healthcare providers.[8]

Measuring pressure invasively, by penetrating the arterial wall to take the measurement, is much less common and usually restricted to a hospital setting.

Non-invasive

The non-invasive auscultatory and oscillometric measurements are simpler and quicker than invasive measurements, require less expertise, have virtually no complications, are less unpleasant and less painful for the patient. However, non-invasive methods may yield somewhat lower accuracy and small systematic differences in numerical results. Non-invasive measurement methods are more commonly used for routine examinations and monitoring. New non-invasive and continuous technologies based on the CNAP vascular unloading technique, are making non-invasive measurement of blood pressure and further advanced hemodynamic parameters more applicable in general anesthesia and surgery where periods of hypotension might be missed by intermittent measurements.[9]

Palpation

A minimum systolic value can be roughly estimated by palpation, most often used in emergency situations, but should be used with caution.[10] It has been estimated that, using 50% percentiles, carotid, femoral and radial pulses are present in patients with a systolic blood pressure > 70 mmHg, carotid and femoral pulses alone in patients with systolic blood pressure of > 50 mmHg, and only a carotid pulse in patients with a systolic blood pressure of > 40 mmHg.[10]

A more accurate value of systolic blood pressure can be obtained with a sphygmomanometer and palpating the radial pulse.[11] Methods using constitutive models have been proposed to measure blood pressure from radial artery pulse.[citation needed] The diastolic blood pressure cannot be estimated by this method. The American Heart Association recommends that palpation be used to get an estimate before using the auscultatory method.

Auscultatory

Measuring systolic and diastolic blood pressure using a mercury sphygmomanometer
Auscultatory method aneroid sphygmomanometer with stethoscope
Mercury manometer

The auscultatory method (from the Latin word for "listening") uses a stethoscope and a sphygmomanometer. This comprises an inflatable (Riva-Rocci) cuff placed around the upper arm at roughly the same vertical height as the heart, attached to a mercury or aneroid manometer. The mercury manometer, considered the gold standard, measures the height of a column of mercury, giving an absolute result without need for calibration and, consequently, not subject to the errors and drift of calibration which affect other methods. The use of mercury manometers is often required in clinical trials and for the clinical measurement of hypertension in high-risk patients, such as pregnant women.

A cuff of the appropriate size[12] is fitted snugly, then inflated manually by repeatedly squeezing a rubber bulb until the artery is completely occluded. It is important that the cuff size is correct: undersized cuffs record too high a pressure; oversized cuffs may yield too low a pressure.[13] Usually three or four cuff sizes should be available to allow measurements in arms of different size.[13] Listening with the stethoscope to the brachial artery at the antecubital area of the elbow, the examiner slowly releases the pressure in the cuff. When blood just starts to flow in the artery, the turbulent flow creates a "whooshing" or pounding (first Korotkoff sound).[14] The pressure at which this sound is first heard is the systolic blood pressure. The cuff pressure is further released until no sound can be heard (fifth Korotkoff sound), at the diastolic arterial pressure.

The auscultatory method is the predominant method of clinical measurement.[15]

Oscillometric

The oscillometric method was first demonstrated in 1876 and involves the observation of oscillations in the sphygmomanometer cuff pressure[16] which are caused by the oscillations of blood flow, i.e., the pulse.[17] The electronic version of this method is sometimes used in long-term measurements and general practice. The first fully automated oscillometric blood pressure cuff called the Dinamap 825, an acronym for "Device for Indirect Non-invasive Mean Arterial Pressure", was made available in 1976.[18] It was replaced in 1978 by the Dinamap 845 which could also measure systolic and diastolic blood pressure, as well as heart rate.[19]

The oscillometric method uses a sphygmomanometer cuff, like the auscultatory method, but with an electronic pressure sensor (transducer) to observe cuff pressure oscillations, electronics to automatically interpret them, and automatic inflation and deflation of the cuff. The pressure sensor should be calibrated periodically to maintain accuracy.[20] Oscillometric measurement requires less skill than the auscultatory technique and may be suitable for use by untrained staff and for automated patient home monitoring. As for the auscultatory technique it is important that the cuff size is appropriate for the arm. There are some single cuff devices that may be used for arms of differing sizes, although experience with these is limited.[13]

The cuff is inflated to a pressure initially in excess of the systolic arterial pressure and then reduced to below diastolic pressure over a period of about 30 seconds. When blood flow is nil (cuff pressure exceeding systolic pressure) or unimpeded (cuff pressure below diastolic pressure), cuff pressure will be essentially constant. When blood flow is present, but restricted, the cuff pressure, which is monitored by the pressure sensor, will vary periodically in synchrony with the cyclic expansion and contraction of the brachial artery, i.e., it will oscillate.

Over the deflation period, the recorded pressure waveform forms a signal known as the cuff deflation curve. A bandpass filter is utilized to extract the oscillometric pulses from the cuff deflation curve. Over the deflation period, the extracted oscillometric pulses form a signal known as the oscillometric waveform (OMW). The amplitude of the oscillometric pulses increases to a maximum and then decreases with further deflation. A variety of analysis algorithms can be employed in order to estimate the systolic, diastolic, and mean arterial pressure.

Oscillometric monitors may produce inaccurate readings in patients with heart and circulation problems, which include arteriosclerosis, arrhythmia, preeclampsia, pulsus alternans, and pulsus paradoxus.[13][21]

In practice the different methods do not give identical results; an algorithm and experimentally obtained coefficients are used to adjust the oscillometric results to give readings which match the auscultatory results as well as possible. Some equipment uses computer-aided analysis of the instantaneous arterial pressure waveform to determine the systolic, mean, and diastolic points. Since many oscillometric devices have not been validated, caution must be given as most are not suitable in clinical and acute care settings.

Recently, several coefficient-free oscillometric algorithms have developed for estimation of blood pressure.[20] These algorithms do not rely on experimentally obtained coefficients and have been shown to provide more accurate and robust estimation of blood pressure. These algorithms are based on finding the fundamental relationship between the oscillometric waveform and the blood pressure using modeling[22] and learning[23] approaches. Pulse transit time measurements have been also used to improve oscillometric blood pressure estimates.[24]

The term NIBP, for non-invasive blood pressure, is often used to describe oscillometric monitoring equipment.

Continuous noninvasive techniques

Continuous Noninvasive Arterial Pressure (CNAP) is the method of measuring beat-to-beat arterial blood pressure in real-time without any interruptions and without cannulating the human body. CNAP combines the advantages of the two clinical gold standards: it measures blood pressure continuously in real-time like the invasive arterial catheter system and it is noninvasive like the standard upper arm sphygmomanometer. Latest developments in this field show promising results in terms of accuracy, ease of use and clinical acceptance. An advanced hemodynamic monitoring system incorporating the CNAP method is the NICCI technology of the company Pulsion Medical Systems [de]. The system uses photoplethysmography to detect the blood flow in the patient's fingers and pressure cuffs to create a constant flow. The resulting pressure in the finger sensor corresponds to the real arterial pressure. Based on the vascular unloading technique, the NICCI Technology provides continuous and noninvasive hemodynamic parameters during surgeries. The measurement results are comparable to invasive arterial line measurements in terms of continuity, accuracy and waveform dynamics.

Recent developments have proposed continuous, noninvasive, non-contact blood pressure measurements using systems such as cameras to monitor the human face.[25]

Pulse wave velocity

Since the 1990s a novel family of techniques based on the so-called pulse wave velocity (PWV) principle have been developed. These techniques rely on the fact that the velocity at which an arterial pressure pulse travels along the arterial tree depends, among others, on the underlying blood pressure.[26] Accordingly, after a calibration maneuver, these techniques provide indirect estimates of blood pressure by translating PWV values into blood pressure values.[27] The main advantage of these techniques is that it is possible to measure PWV values of a subject continuously (beat-by-beat), without medical supervision, and without the need of continuously inflating brachial cuffs.[28]

Ambulatory and home monitoring

Ambulatory blood pressure devices take readings regularly (e.g. every half-hour throughout the day and night). They have been used to exclude measurement problems like white-coat hypertension and provide more reliable estimates of usual blood pressure and cardiovascular risk. Blood pressure readings outside of a clinical setting are usually slightly lower in the majority of people; however studies that quantified the risks from hypertension and the benefits of lowering blood pressure have mostly been based on readings in a clinical environment. Use of ambulatory measurements is not widespread but guidelines developed by the UK National Institute for Health and Care Excellence and the British Hypertension Society recommended that 24-hour ambulatory blood pressure monitoring should be used for diagnosis of hypertension.[29] Health economic analysis suggested that this approach would be cost effective compared with repeated clinic measurements.[30] Not all home blood pressure machines are accurate,[31] and "wide range" (one-size fits all) home blood pressure monitoring units do not have adequate evidence to support their use.[32] In addition, health care professionals are recommending that people validate their home devices before relying on the results.[33]

Home monitoring is a cheap and simple alternative to ambulatory blood pressure monitoring, although it does not usually allow assessment of blood pressure during sleep which may be a disadvantage.[34][35] Automatic self-contained blood pressure monitors are available at reasonable prices, however measurements may not be accurate in patients with atrial fibrillation or other arrhythmias such as frequent ectopic beats.[34][35] Home monitoring may be used to improve hypertension management and to monitor the effects of lifestyle changes and medication related to blood pressure.[36] Compared to ambulatory blood pressure measurements, home monitoring has been found to be an effective and lower cost alternative,[34][37][38] but ambulatory monitoring is more accurate than both clinic and home monitoring in diagnosing hypertension.

When measuring blood pressure in the home, an accurate reading requires that one not drink coffee, smoke cigarettes, or engage in strenuous exercise for 30 minutes before taking the reading. A full bladder may have a small effect on blood pressure readings; if the urge to urinate arises, one should do so before the reading. For 5 minutes before the reading, one should sit upright in a chair with one's feet flat on the floor and with limbs uncrossed. The blood pressure cuff should always be against bare skin, as readings taken over a shirt sleeve are less accurate. The same arm should be used for all measurements. During the reading, the arm that is used should be relaxed and kept at heart level, for example by resting it on a table.[39]

Since blood pressure varies throughout the day, home measurements should be taken at the same time of day. A Joint Scientific Statement From the American Heart Association, American Society of Hypertension, and Preventive Cardiovascular Nurses Association on home monitoring in 2008[35] recommended that 2 to 3 readings should be taken in the morning (after awakening, before washing/dressing, taking breakfast/drink or taking medication) and another 2 to 3 readings at night, each day over a period of 1 week. It was also recommended that the readings from the first day should be discarded and that a total of ≥12 readings (i.e. at least two readings per day for the remaining 6 days of the week) should be used for making clinical decisions.

Observer error

There are many factors that can play a role in the blood pressure reading by physician, such as hearing problem, auditory perception of the physician. Karimi Hosseini et al. evaluated the interobserver differences among specialists without any auditory impairment, and reported 68% of observers recorded systolic blood pressure in a range of 9.4 mmHg, diastolic blood pressure in a range of 20.5 mmHg and mean blood pressure in a range of 16.1mmHg.[citation needed] Neufeld et al. reported standard deviations for both systolic and diastolic readings were roughly 3.5 to 5.5 mm Hg. In general standard deviation for the diastolic pressure would be greater because of the difficulty in judging when the sounds disappear.[40]

White-coat hypertension

For some patients, blood pressure measurements taken in a doctor's office may not correctly characterize their typical blood pressure.[41] In up to 25% of patients, the office measurement is higher than their typical blood pressure. This type of error is called white-coat hypertension (WCH) and can result from anxiety related to an examination by a health care professional.[42] White coat hypertension can also occur because, in a clinical setting, patients are seldom given the opportunity to rest for five minutes before blood pressure readings are taken. The misdiagnosis of hypertension for these patients can result in needless and possibly harmful medication. WCH can be reduced (but not eliminated) with automated blood pressure measurements over 15 to 20 minutes in a quiet part of the office or clinic.[43] In some cases a lower blood pressure reading occurs at the doctor's – this has been termed 'masked hypertension'.[44]

Alternative settings, such as pharmacies, have been proposed as alternatives to office blood pressure monitoring.[45]

Invasive

Arterial blood pressure is most accurately measured invasively through an arterial line. Invasive arterial pressure measurement with intravascular cannulae involves direct measurement of arterial pressure by placing a cannula needle in an artery (usually radial, femoral, dorsalis pedis or brachial). The cannula is inserted either via palpation or with the use of ultrasound guidance.[46]

The cannula must be connected to a sterile, fluid-filled system, which is connected to an electronic pressure transducer. The advantage of this system is that pressure is constantly monitored beat-by-beat, and a waveform (a graph of pressure against time) can be displayed. This invasive technique is regularly employed in human and veterinary intensive care medicine, anesthesiology, and for research purposes.

Cannulation for invasive vascular pressure monitoring is infrequently associated with complications such as thrombosis, infection, and bleeding. Patients with invasive arterial monitoring require very close supervision, as there is a danger of severe bleeding if the line becomes disconnected. It is generally reserved for patients where rapid variations in arterial pressure are anticipated.

Invasive vascular pressure monitors are pressure monitoring systems designed to acquire pressure information for display and processing. There are a variety of invasive vascular pressure monitors for trauma, critical care, and operating room applications. These include single pressure, dual pressure, and multi-parameter (i.e. pressure / temperature). The monitors can be used for measurement and follow-up of arterial, central venous, pulmonary arterial, left atrial, right atrial, femoral arterial, umbilical venous, umbilical arterial, and intracranial pressures.

Central aortic blood pressure measurement

Central aortic blood pressure, which, compared to peripheral blood pressure, has been shown to correlate more strongly with negative cardiovascular events, kidney disease, and other morbidities, as well as all-cause mortality, has traditionally required invasive heart catheterization to be measured, but newer devices have either been developed or are in late stages of development that allow it to be non-invasively measured indirectly with an acceptable margin of error.[47][48][49]

References

  1. ^ Booth J (1977). "A short history of blood pressure measurement". Proceedings of the Royal Society of Medicine. 70 (11): 793–799. doi:10.1177/003591577707001112. PMC 1543468. PMID 341169.
  2. ^ Homan TD, Bordes SJ, Cichowski E (12 July 2022). "Physiology, Pulse Pressure". StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing. PMID 29494015. Retrieved 2019-07-21 – via NCBI Bookshelf.
  3. ^ Franklin, Stanley S.; Khan, Shehzad A.; Wong, Nathan D.; Larson, Martin G.; Levy, Daniel (27 Jul 1999). "Is Pulse Pressure Useful in Predicting Risk for Coronary Heart Disease?". Circulation. 100 (4). Ovid Technologies (Wolters Kluwer Health): 354–360. doi:10.1161/01.cir.100.4.354. ISSN 0009-7322. PMID 10421594.
  4. ^ Benetos, Athanase; Safar, Michel; Rudnichi, Annie; Smulyan, Harold; Richard, Jacques-Lucien; Ducimetière, Pierre; Guize, Louis (1997). "Pulse Pressure". Hypertension. 30 (6). Ovid Technologies (Wolters Kluwer Health): 1410–1415. doi:10.1161/01.hyp.30.6.1410. ISSN 0194-911X. PMID 9403561.
  5. ^ Khilnani P, Singhi S, Lodha R, Santhanam I, Sachdev A, Chugh K, Jaishree M, Ranjit S, Ramachandran B, Ali U, Udani S, Uttam R, Deopujari S (January 2010). "Pediatric Sepsis Guidelines: Summary for resource-limited countries". Indian J Crit Care Med. 14 (1): 41–52. doi:10.4103/0972-5229.63029. PMC 2888329. PMID 20606908.
  6. ^ Zheng L, Sun Z, Li J, Zhang R, Zhang X, Liu S, et al. (July 2008). "Pulse pressure and mean arterial pressure in relation to ischemic stroke among patients with uncontrolled hypertension in rural areas of China". Stroke. 39 (7): 1932–1937. doi:10.1161/STROKEAHA.107.510677. PMID 18451345.
  7. ^ "Blood pressure measurement: MedlinePlus Medical Encyclopedia". medlineplus.gov. Retrieved 2020-04-30.
  8. ^ "Vital Signs (Body Temperature, Pulse Rate, Respiration Rate, Blood Pressure)". OHSU Health Information. Oregon Health & Science University. Retrieved 2014-04-16.
  9. ^ Solà, Josep; Proença, Martin; Braun, Fabian; Pierrel, Nicolas; Degiorgis, Yan; Verjus, Christophe; Lemay, Mathieu; Bertschi, Mattia; Schoettker, Patrick (1 September 2016). "Continuous non-invasive monitoring of blood pressure in the operating room: a cuffless optical technology at the fingertip". Current Directions in Biomedical Engineering. 2 (1): 267–271. doi:10.1515/cdbme-2016-0060.
  10. ^ a b Deakin CD, Low JL (September 2000). "Accuracy of the advanced trauma life support guidelines for predicting systolic blood pressure using carotid, femoral, and radial pulses: observational study". BMJ. 321 (7262): 673–674. doi:10.1136/bmj.321.7262.673. PMC 27481. PMID 10987771.
  11. ^ Interpretation – Blood Pressure – Vitals Archived 2012-07-02 at archive.today, University of Florida, accessed 2008-03-18
  12. ^ Chiappa, M. M.; Ostchega, Y. (2013). "Mean mid-arm circumference and blood pressure cuff sizes for U.S. adults: National Health and Nutrition Examination Survey, 1999-2010". Blood Pressure Monitoring. 18 (3): 138–143. doi:10.1097/MBP.0b013e3283617606. PMC 9681024. PMID 23604196. S2CID 45820114.
  13. ^ a b c d O'Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G, Mengden T, Myers M, Padfield P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P (2003). "European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement". J. Hypertens. 21 (5): 821–848. doi:10.1097/00004872-200305000-00001. PMID 12714851. S2CID 3952069.
  14. ^ Audio recordings of Korotkoff sounds. Slide seven of twenty-two, Cardiff University/Prfysgol Caerdydd. Accessed 27 September 2016.
  15. ^ (Pickering et al. 2005, p. 146) See Blood Pressure Measurement Methods.
  16. ^ (Pickering et al. 2005, p. 147) See The Oscillometric Technique.
  17. ^ Laurent, P. (2003-09-28). "Blood Pressure & Hypertension". Retrieved 2009-10-05.
  18. ^ "Dinamap 1045". The Wood Library-Museum. Retrieved 14 December 2020.
  19. ^ Westhorpe, R. N. (May 2009). "Blood Pressure Monitoring – Automated Non-invasive Blood Pressure Monitors". Anaesthesia and Intensive Care. 37 (3): 343. doi:10.1177/0310057X0903700321. PMID 19499853.
  20. ^ a b Forouzanfar, M.; Dajani, H. R.; Groza, V. Z.; Bolic, M.; Rajan, S.; Batkin, I. (2015-01-01). "Oscillometric Blood Pressure Estimation: Past, Present, and Future". IEEE Reviews in Biomedical Engineering. 8: 44–63. doi:10.1109/RBME.2015.2434215. ISSN 1937-3333. PMID 25993705. S2CID 8940215.
  21. ^ Hamzaoui O, Monnet X, Teboul JL (2013). "Pulsus paradoxus". Eur. Respir. J. 42 (6): 1696–1705. doi:10.1183/09031936.00138912. PMID 23222878. S2CID 35428633.
  22. ^ Forouzanfar M (2014). "Ratio-independent blood pressure estimation by modeling the oscillometric waveform envelope". IEEE Trans. Instrum. Meas. 63 (10): 2501–2503. Bibcode:2014ITIM...63.2501F. doi:10.1109/tim.2014.2332239. S2CID 7494219.
  23. ^ Forouzanfar, M.; Dajani, H. R.; Groza, V. Z.; Bolic, M.; Rajan, S. (August 2011). "Feature-Based Neural Network Approach for Oscillometric Blood Pressure Estimation". IEEE Transactions on Instrumentation and Measurement. 60 (8): 2786–2796. Bibcode:2011ITIM...60.2786F. doi:10.1109/TIM.2011.2123210. S2CID 7614742.
  24. ^ Forouzanfar, M.; Ahmad, S.; Batkin, I.; Dajani, H. R.; Groza, V. Z.; Bolic, M. (2013-07-01). "Coefficient-Free Blood Pressure Estimation Based on Pulse Transit Time #x2013;Cuff Pressure Dependence". IEEE Transactions on Biomedical Engineering. 60 (7): 1814–1824. doi:10.1109/TBME.2013.2243148. ISSN 0018-9294. PMID 23372068. S2CID 1578853.
  25. ^ Iuchi, Kaito (2022). "Blood pressure estimation by spatial pulse-wave dynamics in a facial video". Biomedical Optics Express. 13 (11): 6035–6047. doi:10.1364/BOE.473166. PMC 9872895. PMID 36733727.
  26. ^ Asmar, Roland (1999). Arterial Stiffness and Pulse Wave Velocity. Paris: Elsevier. ISBN 978-2-84299-148-7.
  27. ^ Solà, Josep (2011). Continuous non-invasive blood pressure estimation (PDF). Zurich: ETHZ PhD dissertation.
  28. ^ Solà, Josep; Delgado-Gonzalo, Ricard (2019). The Handbook of Cuffless Blood Pressure Monitoring. Springer International Publishing. ISBN 978-3-030-24701-0.
  29. ^ Hypertension Guideline 2011 [CG127] Produced in a collaboration between the British Hypertension Society and NICE. http://guidance.nice.org.uk/CG127/NICEGuidance/pdf/English
  30. ^ Lovibond K, Jowett S, Barton P, Caulfield M, Heneghan C, Hobbs FD, Hodgkinson J, Mant J, Martin U, Williams B, Wonderling D, McManus RJ (2011). "Cost-effectiveness of options for the diagnosis of high blood pressure in primary care: a modelling study". Lancet. 378 (9798): 1219–1230. doi:10.1016/S0140-6736(11)61184-7. PMID 21868086. S2CID 5151024.
  31. ^ O'Brien, Eoin; Stergiou, George S.; Turner, Martin J. (2018). "The quest for accuracy of blood pressure measuring devices". Journal of Clinical Hypertension. 20 (7): 1092–1095. doi:10.1111/jch.13279. ISSN 1751-7176. PMC 8031088. PMID 30003703.
  32. ^ Sprague, Eliott; Padwal, Raj S. (2018). "Adequacy of validation of wide-range cuffs used with home blood pressure monitors: a systematic review". Blood Pressure Monitoring. 23 (5): 219–224. doi:10.1097/MBP.0000000000000344. ISSN 1473-5725. PMID 30074520. S2CID 51908913.
  33. ^ Ruzicka, Marcel; Hiremath, Swapnil (2017-07-01). "Accuracy-Limiting Factor of Home Blood Pressure Monitors?". American Journal of Hypertension. 30 (7): 661–664. doi:10.1093/ajh/hpx056. ISSN 1941-7225. PMID 28430845.
  34. ^ a b c Mancia G, De Backer G, Dominiczak A, Cifkova R, Fagard R, Germano G, Grassi G, Heagerty AM, Kjeldsen SE, Laurent S, Narkiewicz K, Ruilope L, Rynkiewicz A, Schmieder RE, Struijker Boudier HA, Zanchetti A, Vahanian A, Camm J, De Caterina R, Dean V, Dickstein K, Filippatos G, Funck-Brentano C, Hellemans I, Kristensen SD, McGregor K, Sechtem U, Silber S, Tendera M, Widimsky P, Zamorano JL, Kjeldsen SE, Erdine S, Narkiewicz K, Kiowski W, Agabiti-Rosei E, Ambrosioni E, Cifkova R, Dominiczak A, Fagard R, Heagerty AM, Laurent S, Lindholm LH, Mancia G, Manolis A, Nilsson PM, Redon J, Schmieder RE, Struijker-Boudier HA, Viigimaa M, Filippatos G, Adamopoulos S, Agabiti-Rosei E, Ambrosioni E, Bertomeu V, Clement D, Erdine S, Farsang C, Gaita D, Kiowski W, Lip G, Mallion JM, Manolis AJ, Nilsson PM, O'Brien E, Ponikowski P, Redon J, Ruschitzka F, Tamargo J, van Zwieten P, Viigimaa M, Waeber B, Williams B, Zamorano JL (June 2007). "2007 Guidelines for the management of arterial hypertension: The Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC)". Eur Heart J. 28 (12): 1462–1536. doi:10.1093/eurheartj/ehm236. PMID 17562668.
  35. ^ a b c Pickering TG, Miller NH, Ogedegbe G, Krakoff LR, Artinian NT, Goff D (2008). "Call to action on use and reimbursement for home blood pressure monitoring: executive summary: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association". Hypertension. 52 (1): 1–9. doi:10.1161/HYPERTENSIONAHA.107.189011. PMID 18497371.
  36. ^ Chobanian AV, Bakris GL, Black HR, Cushman WC, Green LA, Izzo JL, Jones DW, Materson BJ, Oparil S, Wright JT, Roccella EJ (December 2003). "Seventh report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure". Hypertension. 42 (6): 1206–1252. doi:10.1161/01.HYP.0000107251.49515.c2. PMID 14656957.
  37. ^ Niiranen TJ, Kantola IM, Vesalainen R, Johansson J, Ruuska MJ (2006). "A comparison of home measurement and ambulatory monitoring of blood pressure in the adjustment of antihypertensive treatment". Am J Hypertens. 19 (5): 468–474. doi:10.1016/j.amjhyper.2005.10.017. PMID 16647616.
  38. ^ Shimbo D, Pickering TG, Spruill TM, Abraham D, Schwartz JE, Gerin W (2007). "The Relative Utility of Home, Ambulatory, and Office Blood Pressures in the Prediction of End-Organ Damage". Am J Hypertens. 20 (5): 476–482. doi:10.1016/j.amjhyper.2006.12.011. PMC 1931502. PMID 17485006.
  39. ^ "How to prepare for a blood pressure test". National Heart, Lung and Blood Institute. Archived from the original on 2021-04-07.
  40. ^ Neufeld, PD; Johnson, DL (15 September 1986). "Observer error in blood pressure measurement". Canadian Medical Association Journal. 135 (6): 633–637. PMC 1491295. PMID 3756693.
  41. ^ Elliot, Victoria Stagg (2007-06-11). "Blood pressure readings often unreliable". American Medical News. American Medical Association. Archived from the original on 2008-11-12. Retrieved 2008-08-16.
  42. ^ Jhalani J, Goyal T, Clemow L, Schwartz JE, Pickering TG, Gerin W (2005). "Anxiety and outcome expectations predict the white-coat effect". Blood Pressure Monitoring. 10 (6): 317–319. doi:10.1097/00126097-200512000-00006. PMID 16496447. S2CID 2058260.
  43. ^ (Pickering et al. 2005, p. 145) See White Coat Hypertension or Isolated Office Hypertension.
  44. ^ (Pickering et al. 2005, p. 146) See Masked Hypertension or Isolated Ambulatory Hypertension.
  45. ^ Albasri, Ali; OʼSullivan, Jack W.; Roberts, Nia W.; Prinjha, Suman; McManus, Richard J.; Sheppard, James P. (October 2017). "A comparison of blood pressure in community pharmacies with ambulatory, home and general practitioner office readings". Journal of Hypertension. 35 (10): 1919–1928. doi:10.1097/HJH.0000000000001443. PMC 5585128. PMID 28594707.
  46. ^ White, L; Halpin, A; Turner, M; Wallace, L (22 April 2016). "Ultrasound-guided radial artery cannulation in adult and paediatric populations: a systematic review and meta-analysis". British Journal of Anaesthesia. 116 (5): 610–617. doi:10.1093/bja/aew097. PMID 27106964.
  47. ^ O’Rourke, Michael F.; Adji, Audrey (15 Nov 2011). "Noninvasive Studies of Central Aortic Pressure". Current Hypertension Reports. 14 (1). Springer Science and Business Media LLC: 8–20. doi:10.1007/s11906-011-0236-5. ISSN 1522-6417. PMID 22083214.
  48. ^ Liu, Wenyan; Du, Shuo; Zhou, Shuran; Mei, Tiemin; Zhang, Yuelan; Sun, Guozhe; Song, Shuang; Xu, Lisheng; Yao, Yudong; Greenwald, Stephen E. (2022). "Noninvasive estimation of aortic pressure waveform based on simplified Kalman filter and dual peripheral artery pressure waveforms". Computer Methods and Programs in Biomedicine. 219. Elsevier BV: 106760. doi:10.1016/j.cmpb.2022.106760. ISSN 0169-2607. PMID 35338889.
  49. ^ Mariscal-Harana, Jorge; Charlton, Peter H.; Vennin, Samuel; Aramburu, Jorge; Florkow, Mateusz Cezary; van Engelen, Arna; Schneider, Torben; de Bliek, Hubrecht; Ruijsink, Bram; Valverde, Israel; Beerbaum, Philipp; Grotenhuis, Heynric; Charakida, Marietta; Chowienczyk, Phil; Sherwin, Spencer J.; Alastruey, Jordi (1 Feb 2021). "Estimating central blood pressure from aortic flow: development and assessment of algorithms". American Journal of Physiology. Heart and Circulatory Physiology. 320 (2). American Physiological Society: H494 – H510. doi:10.1152/ajpheart.00241.2020. hdl:10261/265391. ISSN 0363-6135. PMC 7612539. PMID 33064563.

Sources

Read other articles:

bra full-cupPlunge Beha, BH, bra atau kutang adalah pakaian dalam dengan bentuk pas yang dirancang untuk menopang atau menutupi payudara wanita. Bra dirancang untuk berbagai tujuan, termasuk memperbesar ukuran payudara, membuat belahan dada, atau untuk pertimbangan estetika atau praktis lainnya. Pakaian renang, kamisol, dan gaun tanpa bagian belakang mungkin memiliki penyangga payudara built-in dengan cup bra yang mendukung. Bra menyusui dirancang untuk memudahkan menyusui. Beberapa orang mem...

 

Artikel ini perlu diwikifikasi agar memenuhi standar kualitas Wikipedia. Anda dapat memberikan bantuan berupa penambahan pranala dalam, atau dengan merapikan tata letak dari artikel ini. Untuk keterangan lebih lanjut, klik [tampil] di bagian kanan. Mengganti markah HTML dengan markah wiki bila dimungkinkan. Tambahkan pranala wiki. Bila dirasa perlu, buatlah pautan ke artikel wiki lainnya dengan cara menambahkan [[ dan ]] pada kata yang bersangkutan (lihat WP:LINK untuk keterangan lebih lanjut...

 

Formal fallacy WhataboutismTacticPropaganda techniqueTypeTu quoque (appeal to hypocrisy)LogicLogical fallacy Whataboutism or whataboutery (as in what about…?) is a pejorative for the strategy of responding to an accusation with a counter-accusation instead of a defense of the original accusation. From a logical and argumentative point of view, whataboutism is considered a variant of the tu-quoque pattern (Latin 'you too', term for a counter-accusation), which is a subtype of the ad-hominem ...

This article needs additional citations for verification. Please help improve this article by adding citations to reliable sources. Unsourced material may be challenged and removed.Find sources: Loveland Block and Coors Building – news · newspapers · books · scholar · JSTOR (February 2011) (Learn how and when to remove this template message) United States historic placeLoveland Building and Coors BuildingU.S. National Register of Historic Places Lovela...

 

Pour les articles homonymes, voir Fibre (homonymie). Fibre optiqueUn faisceau de fibres optiques.Type Fibre de verre, guide d'onde optique (en)CaractéristiquesComposé de Guide d'onde optique (en)modifier - modifier le code - modifier Wikidata Une fibre optique est un fil dont l’âme, très fine et faite de verre ou de plastique, a la propriété de conduire la lumière et sert pour la fibroscopie, l'éclairage ou la transmission de données numériques. Elle offre un débit d'information...

 

Bagian dari seriPendidikan di Indonesia Kementerian Pendidikan, Kebudayaan, Riset, dan Teknologi Republik Indonesia Pendidikan anak usia dini TK RA KB Pendidikan dasar (kelas 1–6) SD MI Paket A Pendidikan dasar (kelas 7–9) SMP MTs Paket B Pendidikan menengah (kelas 10–12) SMA MA SMK MAK SMA SMTK SMAK Utama Widya Pasraman Paket C Pendidikan tinggi Perguruan tinggi Akademi Akademi komunitas Institut Politeknik Sekolah tinggi Universitas Lain-lain Madrasah Pesantren Sekolah alam Sekolah ru...

Untuk kegunaan lain, lihat Kaki (disambiguasi). kaki Kaki merupakan salah satu anggota tubuh hewan atau manusia yang digunakan untuk berjalan, berlari atau melompat. Kaki terdiri dari beberapa bagian, termasuk telapak kaki, sendi yang bekerja dalam suatu sistem terpadu sehingga memungkinkan bagi inang untuk berjalan. Referensi France, Diane L. (2008). Human and Nonhuman Bone Identification: A Color Atlas. CRC Press. ISBN 1-4200-6286-7.  Marieb, Elaine Nicpon; Hoehn, Katja (2007). Hu...

 

Pour les articles homonymes, voir George Herbert Walker (homonymie), George Walker, Walker, George Bush et Bush. Pour les autres membres de la famille, voir Famille Bush. Ne pas confondre avec George W. Bush, son fils. George H. W. Bush Portrait officiel de George H. W. Bush (1981). Fonctions 41e président des États-Unis 20 janvier 1989 – 20 janvier 1993(4 ans) Élection 8 novembre 1988 Vice-président Dan Quayle Gouvernement Administration George H. W. Bush Prédécesseur Ronald Re...

 

Questa voce sull'argomento cardinali italiani è solo un abbozzo. Contribuisci a migliorarla secondo le convenzioni di Wikipedia. Ferdinando Maria Saluzzocardinale di Santa Romana ChiesaRitratto del cardinale Saluzzo, tra 1801 e 1806, presso il Santuario della Madonna di Schiavonea a Corigliano Calabro  Incarichi ricoperti Arcivescovo titolare di Teodosia (1784) Arcivescovo titolare di Cartagine (1784-1801) Nunzio apostolico in Polonia (1784-1794) Presidente della Legazione di Urbi...

У этого термина существуют и другие значения, см. Ухо (значения). Анатомия уха. Наружное ухо: 1 — височная кость 2 — слуховой канал 3 — ушная раковина среднее ухо: 4 — барабанная перепонка 6 — молоточек 7 — наковальня 8 — стремечко внутре...

 

Pour les articles homonymes, voir Cloutier. Véronique CloutierVéronique Cloutier et son mari, l'humoriste et producteur Louis Morissette, au gala Les Olivier.BiographieNaissance 31 décembre 1974 (49 ans)LavalNationalité canadienneFormation Collège Jean-de-BrébeufActivités Actrice, productrice, humoriste, animatrice de radio, animatrice de télévision, femme d'affairesPériode d'activité depuis 1990Père Guy CloutierFratrie Stéphanie Cloutier (d)Conjoint Louis MorissetteAutres ...

 

PlanocityCity of Plano Plano – VedutaIl Granite Park nel maggio del 2011 LocalizzazioneStato Stati Uniti Stato federato Texas ConteaCollinDenton AmministrazioneSindacoJohn B. Muns (I) TerritorioCoordinate33°03′N 96°45′W / 33.05°N 96.75°W33.05; -96.75 (Plano)Coordinate: 33°03′N 96°45′W / 33.05°N 96.75°W33.05; -96.75 (Plano) Altitudine206 m s.l.m. Superficie186,33 km² Abitanti259 841 (2010) Densità1 3...

American politician from Connecticut Lucy Dathan is an American politician who serves as a member of the Connecticut House of Representatives from the 142nd district, comprising parts of New Canaan and Norwalk.[1] She was born in St. Louis, Missouri, and has three children.[2] Career Dathan, a Democrat affiliated with the Working Families Party, ousted incumbent Republican Fred Wilms in the 2018 Connecticut elections to become the first Democrat to win election in the 142nd Ho...

 

Ця стаття потребує додаткових посилань на джерела для поліпшення її перевірності. Будь ласка, допоможіть удосконалити цю статтю, додавши посилання на надійні (авторитетні) джерела. Зверніться на сторінку обговорення за поясненнями та допоможіть виправити недоліки. Мат...

 

Translations of the Bible This article is about the general history of Bible translations. For translations into specific languages, see List of Bible translations by language. A selection of Bible translations in contemporary English. Part of a series on theBible Canons and books Tanakh Torah Nevi'im Ketuvim Old Testament (OT) New Testament (NT) Deuterocanon Antilegomena Chapters and verses Apocrypha Jewish OT NT Authorship and development Authorship Dating Hebrew canon Old Testament canon N...

Nero's Weekend-Mio figlio Nerone-SutradaraStenoProduserFranco CristaldiDitulis olehStefano Vanzina, Alberto Sordi, Rodolfo Sonego, Sandro ContinenzaPemeranAlberto SordiVittorio De SicaGloria SwansonBrigitte BardotPenata musikAngelo Francesco LavagninoSinematograferMario BavaPenyuntingMario SerandreiPerusahaanproduksiCinecittàDistributorTitanusTanggal rilis 1956 (1956) Durasi95 menitNegaraItaliaBahasaItaliaPendapatankotor571.983 pemasukan (Prancis)[1] Brigitte Bardot, Vitto...

 

World's fair held in Seville, Spain This article includes a list of general references, but it lacks sufficient corresponding inline citations. Please help to improve this article by introducing more precise citations. (April 2016) (Learn how and when to remove this message) 1929 SevillePosterOverviewBIE-classUniversal expositionCategoryHistorical ExpoNameExposición General de España (section:Exposición Ibero-Americana)Area69 hectares (170 acres)Participant(s)Countries18LocationCountrySpai...

 

French soldier (1661–1706) Pierre Le Moyne d'IbervilleNational Library and Archives of QuebecBorn(1661-07-16)16 July 1661Ville-Marie, New FranceDied9 July 1706(1706-07-09) (aged 44)Havana, Spanish CubaAllegiance Kingdom of FranceYears of service1686–1706Battles/warsKing William's War Hudson Bay Campaign Battle of Fort Albany Raid on Schenectady Capture of York Factory Action of 14 July 1696 Siege of Pemaquid Avalon Peninsula Campaign Battle of Hudson's Bay War of the Spanis...

Artikel ini tidak lengkap. Silakan bantu memperbaiki artikel ini, atau diskusikan hal ini di halaman pembicaraan. Pos Indonesia merupakan perusahaan tertua di Indonesia menurut awal sejarah yang diakui perusahaan. Berikut ini daftar perusahaan tertua di Indonesia yang didirikan sebelum tahun 1945 dan dihitung menurut pengakuan perusahaan saat ini atau anggapan masyarakat pada umumnya. Daftar berikut mencakup perusahaan dan merek, namun tidak mencakup pabrik gula atau organisasi asosiasi, pend...

 

Federasi Sepak Bola Arab SaudiAFCDidirikan1959Kantor pusatRiyadhBergabung dengan FIFA1959Bergabung dengan AFC1959PresidenSultan Bin Abdul AzizWebsitehttps://www.saff.com.sa/en/ Federasi Sepak Bola Arab Saudi (bahasa Arab: الاتحاد السعودي لكرة القدم) merupakan badan sepak bola Arab Saudi. Badan ini didirikan pada 1959, dan bergabung baik dengan FIFA maupun dengan AFC pada tahun yang sama. Badan ini bertanggung jawab pada administrasi kompetisi antar klub di Arab Saud...